All Categories > By Primary Gas to be Measured > Oxygen Analyzers (O2) > Continuous Oxygen Analyzers (O2) > 970 Series Syngas > Item # 975A  
Download PDF Download PDF     Printable Page     Email Email This Page     Save To Favorites Save To Favorites

Item # 975A, Continuous Analyzer for Syn-Gas Atmospheres - O2, CO, CO2, CH4 and H2



The Nova 970 Series Syn-Gas Analyzer System has been designed with the flexibility and robustness required in the experimental and developmental environments that characterize syn gas applications.

The Nova 970 Series utilizes high-stability infrared detectors for the simultaneous measurement of CO, CO2, and CH4. In addition, the analyzer can also be supplied with a non-consumable, long-life thermal conductivity cell for H2 that compensates for the interference effects of CO, CO2, and CH4. This ensures that H2 will always read correctly regardless of the background gas composition.  A non-depleting paramagnetic sensor is used for O2 analysis.  All sensors/detectors are temperature-controlled or temperature-compensated for maximum analytical stability.  Auto calibration functionality allows easy calibration without user intervention.  The included Serial Output Package allows connection of analyzer to personal computers via a choice of USB/RS-232/RS-485. Data recording software is also included.






Specifications  · Applications  · Features  · Options  · Calibration

Specifications

Method of Detection

Electrochemical sensor for O2
NDIR infrared detector for CO, CO2 and CH4

Thermal conductivity cell for H2

Ranges Available

0-10.0%, 0-50.0%, 0-100.0% CO
0-10.0%, 0-50.0%, 0-100.0% CO2

0-2.0%, 0-25.0%, 0-50.0%, 0-100.0% O2

0-5.0%, 0-50.0%, 0-100.0% H2

0-50.0%, 0-100.0% CH4

Resolution

0.1% for all gases

Accuracy & Repeatability

± 1.5% of full scale on all gases

Drift

Less than 2% of full scale per month

Response Time (T-90)

20-30 seconds to 90% step change

Ambient Temperature Range

40-104 °F (4 to 40°C). Optional -22 to 158 °F (-30 to 70 °C) with Outdoor Packages

Linearity

± 1.5% of full scale

Size & Weight

24" H x 24" W x 10" D @ 90 lbs (61 x 61 x 25 cm @ 20 kg)

Power

115 VAC 60 Hz (220 VAC 50 Hz available)

Output Options

Isolated 4-20 mA standard
RS232, RS485, MODBUS®, Ethernet outputs optional

Alarms

High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating.


Applications

For continuous monitoring of syngas and gasification atmospheres or other industrial process gases for any combination of methane (CH4), carbon dioxide (CO2), carbon monoxide (CO) and hydrogen (H2) and oxygen (O 2).



Features

  • Infrared detector for reliable measurement of CO, CO2 and CH4. Detector may be cleaned in the field.
  • Thermal conductivity cell for H2
  • Oxygen analysis by choise of electrochemical or paramagnetic oxygen sensor
  • Sensors/detectors temperature controlled or compensated for maximum analytical stability
  • H2 reading is compensated for the interference effects of the other gases measured
  • High ranges on each channel available (up to 100%)
  • Touch-screen display for gas readings
  • Built-in sample pump or pressure regulator
  • Automatic calibration & serial output package
  • Isolated 4-20 mA analog recorder outputs
  • Durable stainless steel lines & sample components (where prudent)
  • Automatic moisture removal system included



  • Options

  • Calculations performed on gas readings such as gas ratios and heating value of flammable gases
  • Hi/Low gas, low flow and other alarms available
  • Sample conditioning systems available for acidic gas and dust
  • Cabinet purge system available for use in hazardous areas
  • Outdoor packages available for operation from -22 to 158 °F (-30 to 70 °C)
  • Cabinet coolers can be fitted to most models
  • Heated filters and high temperature probes
  • Methane-specific detector in place of standard hydrocarbons detector; allows more accurate CH4 analysis in other HC's



  • Calibration

  • Ambient air for O2 and to zero all other gases
  • Analyzed calibration gas with representative concentrations for span of all other gases



  •  DOWNLOADS 

    970 Series Syn-Gas Analyzer
    (PDF, 576KB)