All Categories > By Primary Gas to be Measured > Oxygen Analyzers (O2) > Continuous Oxygen Analyzers for Hazardous Locations (O2) > Continuous Process Oxygen Gas Analyzer by Paramagnetic Sensor - 412 Series > Item # Model 412N7MC  
Download PDF Download PDF     Printable Page     Email Email This Page     Save To Favorites Save To Favorites

Item # Model 412N7MC, Model 412N7MC Paramagnetic Oxygen Analyzer, Wall Mounted NEMA7 UL/CSA Explosion-proof with Non-intrusive Magnetic Calibration, Class 1 Div 1 Group BCD



The Nova FLEXI-NOVA Series Platform has been designed for continuous gas measurements in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series. Depending on application, some light-duty sample conditioning features may be available.

For clean process measurements, percent-level O2 gas. This sensor utilizes a paramagnetic measuring cell and powerful selenium cobalt magnet assembly to make use of the principle that oxygen is drawn into a magnetic field, thereby offering a method of detection that is very specific to oxygen. Sample dew point should be at least 5°C (9°F) below lowest sample temperature.

All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls.

Cabinet: NEMA 7 ex-proof cabinet, Class 1 Div 1 Group BCD





Specifications  · Applications  · Features  · Options  · Calibration  · Unique Applications

Specifications

Method of Detection

Magneto-dynamic paramagnetic O2 cell

Ranges Available

With back-pressure control: 1.00, 2.00, 5.00, 25.00, 100.0 % O2
Without back-pressure control: 1.0, 2.0, 5.0 % O2

Resolution

0.1% or 0.01% O2, depending on back-pressure control

Accuracy & Repeatability

±1% of full scale

Drift

±0.2% of full scale per month

Response Time (T-90)

4 - 5 seconds at a sample flow of 1 LPM

Ambient Temperature Range

32 to 122°F (0 to 50°C) @ 5-95% RH non-condensing. Lower temperatures (-5°F, -20°C) with Cold Weather Package. Sample dew point should be at least 9°F (5°C) below lowest sample temperature.

Linearity

±0.5% of full scale or 0.05% absolute, whichever is greater, depending on back-pressure control

Size & Weight

EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D)

Power

115 VAC, 60 Hz
220 VAC, 50 Hz

Output Options

4-20mA into 500 ohms non-isolated standard

Alarms

High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating

Analyzer Type

O2 Analyzer

Hazardous Area Rating Options

Class 1, Div 1, Grp A-D Hazardous Area Rating


Applications
      For continuous analysis of oxygen (O2) in process gas streams that contain gases that interfere with other methods of O2 measurement. For high accuracy or high purity applications.




Features
  • Highest accuracy O2 analyzer, very stable
  • Fast response (T90 - 4 to 5 seconds) and bright digital readout
  • Linear over entire 0-100% range
  • Microprocessor control
  • Non-consumable O2 detector, no periodic replacement required
  • Built in sample pump or pressure regulator, and 4-20mA output




Options
  • Hi/Low gas, low flow alarms available




Calibration
  • Nitrogen for zero
  • Air or analyzed calibration gas for span




Unique Applications
    All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.




 DOWNLOADS 

412 FlexiNova O2 by Paramagnetic Analyzer21
(PDF, 606KB)

FlexiNova Ex-Proof Platform Analyzer21
(PDF, 499KB)