Home
About Us
About NOVA
What's new at NOVA
About Tenova Goodfellow
About the Tenova Group
Tenova Group Website
Online Catalog
Gas Analyzers
O
2
Analyzers
CO Analyzers
CO
2
Analyzers
H
2
Analyzers
CH
4
Analyzers
NO
x
Analyzers
NO
2
Analyzers
SO
2
Analyzers
LEL Analyzers
Dew point Monitors
HC's Analyzers
H
2
S Analyzers
NH
3
Analyzers
He Analyzers
SF
6
Analyzers
Multigas
Applications
Flue Gas Analyzers
Emissions Analyzers
Process Gas Analyzers
Landfill Gas Analyzers
Biogas Analyzers
Handheld Landfill & Biogas
Heat Treating Analyzers
Syngas Production Analyzers
Steel-making Analyzers
Hydrogen Purity Analyzers
Ambient Air Monitors
Personnel Safety Monitors
Engine Exhaust Analyzers
Controlled Atmosphere Analyzers
Customer Service
Quick Question
FAQs
Spare Parts
Service Department
Application Questionnaire
Representatives
Site Map
Request Quote
Contact Us
Online Catalog
View Our Online Catalog.
Or search our online
catalog by entering in a
keyword or product
number in the box.
By Model
By Keyword
Applications
Flue Gas Analyzers.
Emissions Analyzers.
Process Gas Analyzers.
Landfill Gas Analyzers.
Biogas Analyzers.
Heat Treating Analyzers.
Syngas Production Analyzers.
Steel-Making Analyzers.
Hydrogen Purity Gas Analyzers.
Ambient Air Monitors.
Personnel Safety Monitors.
Engine Exhaust Analyzers.
Controlled Atmosphere Analyzers.
All Categories
>
By Primary Gas to be Measured
>
Oxygen Analyzers (O
2
)
>
Portable Oxygen Analyzers (O2)
>
Portable Flue Gas Oxygen Analyzer - 320 Series
>
Item # Model 320K
Download PDF
Download PDF
Printable Page
Email This Page
Save To Favorites
Please wait…
Item # Model 320K, Portable Flue Gas Analyzer for Oxygen, Suitcase (K) Enclosure
larger image
The Nova 320 Series of Portable Flue Gas Oxygen Analyzers utilize a reliable, stable oxygen sensor which responds quickly to the oxygen present in a flue gas sample. Its life expectancy is between 3 and 4 years. The sensor is very easy for the customer to replace and at a reasonable cost.
A rechargeable battery provides enough power for the pump and electronics for about 8 hours of operation. The analyzer can still be operated while it is being charged. The analyzer is supplied with recharger and a 12ft (4m) sample hose and a stainless steel probe.
The Model 320 can optionally indicate stack temperature (Model 320T; 0-1800°F or 0-1000°C) for doing fuel efficiency calculations. The temperature sensor is built into the sampling probe. Efficiency charts for each fuel are provided.
Specifications
·
Applications
·
Features
·
Options
·
Calibration
·
Unique Applications
Specifications
Method of Detection
Long life electrochemical oxygen sensor
Ranges Available
0-5.0%, 0-10.0%, 0-25.0%, 0-50.0%
Resolution
0.1% O
2
Accuracy & Repeatability
±1% of full scale (±2% of full scale above 30% O
2
)
Drift
< 2% of full scale for 8 hours of operation
Response Time (T-90)
8-10 seconds for 90% of step change
Ambient Temperature Range
0 to 40
ºC
32 to 105
ºF
Linearity
±1% of full scale
Size & Weight
approx. 26 x 18 x 13 cm @ 3.4 kg (10" x 7" x 5" @ 7.5lbs)
Power
12V battery operation. 115VAC 60Hz for recharging (220VAC 50Hz available)
Output Options
4-20mA
Hazardous Area Rating Options
General Purpose, Non-Hazardous
Applications
Analysis of oxygen (O
2
) in flue gases from any type of fuel. For checking combustion efficiency, air infiltration and burner & control performance of furnaces, heaters and boilers. May be used in commercial, industrial and residential settings.
Features
Rugged design that is easy to operate and maintain
Fast warm-up and response
Long-life electrochemical oxygen sensor (3-4 year life)
Digital readout meter
Rechargeable battery operation
Built in sample pump, filter and flow meter
Continuous condensate removal
For use on any fuel
Suitcase cabinet (K) weatherproof when closed
Stainless steel probe with sample hose
Pays for itself in months through fuel savings
Options
Output - 4-20mA
Special sampling probes
Sample pre-cooler
Suitcase (K) style cabinet available
On-board data-logging to USB stick
Calibration
On air for span
On nitrogen for zero
Unique Applications
All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.
MORE IMAGES
·
Application flue gas
DOWNLOADS
320 Portable O2 Flue Gas Analyzer
(PDF, 608KB)