Item # Model 207K-KX, Model 207K-KX, Portable Standard Range Dew Point Analyzer



The Nova 207 Series Portable Dew Point Analyzers have been designed primarily for accurate dew point measurement of exothermic and endothermic heat treating atmospheres.

The dielectric polymer sensor provides stable and accurate dew point measurements and is not affected by most gas compositions. (some exceptions)

A built-in sample pump draws in samples that are of insufficient pressure from the process. The sample flow is shown on a small rotameter with valve. The analyzer then reads the dew point continuously.



Models:
  • 207K - suitcase cabinet with display on inside; display visible when case is opened
  • 207KX - suitcase cabinet with display on outside; display visible when case is closed





Specifications  · Applications  · Features  · Options  · Calibration  · Unique Applications

Specifications

Method of Detection

Dielectric polymer sensor that change capacitance with moisture variation

Ranges Available

-40°C to +40°C (or -40°F to +104°F) Dew Point

Resolution

1°F (1°C) Dew Point

Accuracy & Repeatability

±1°F (±1°C) Dew Point

Drift

Less than 1% of full scale range per month

Response Time (T-90)

Less than 30 seconds

Ambient Temperature Range

4° to 49°C ºC
40° to 120°F
ºF

Linearity

Better than 1°F (1°C) Dew Point

Size & Weight

K Suitcase Style - Approx. 14" L x 6" H x 11" D @ 12 lbs (36 x 15 x 27 cm @ 5.4 kg

Power

115 VAC 60 Hz for recharging
220VAC 50Hz available

Output Options

4-20mA

Alarms

High or low dew point alarms with LED (optional)

Hazardous Area Rating Options

General Purpose, Non-Hazardous


Applications
For checking dew points of air, heat treat furnace gases, compressed air lines, and other applications with oils or contaminants present.



Features
  • Standard range of -40 to +104°F (-40 to +40°C) Dew Point
  • Rugged design that is easy to operate
  • Use on any non-corrosive gas
  • Long-life dielectric polymer sensor
  • May be calibrated in the field
  • Bright digital readout
  • Built in sample pump, filter and flow meter
  • Suitcase style (K) enclosure is standard
  • Sensor easily replaced if contaminated
  • Factory calibrated to NIST traceable standards




Options
  • 4-20mA recorder output
  • Calibration kit
  • Built-in printer
  • (KX) display on outside of case




Calibration
  • Calibration is easily performed in the field using the optional calibration kit (part no. TGI50317SUB).
  • Alternatively, the analyzer may be returned to Nova for a certified calibration




Unique Applications
The 207 Series should not be used to measure dew points when trace amounts of sulfur trioxide (SO3), chlorine (Cl2), hydrogen chloride (HCl), or ammonia (NH3) are present. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.