All Categories > By Primary Gas to be Measured > Oxygen Analyzers (O2) > Continuous Oxygen Analyzers (O2) > Continuous Process Oxygen Gas Analyzer - 410 Series > Item # Model 410N7MC  
Download PDF Download PDF     Printable Page     Email Email This Page     Save To Favorites Save To Favorites

Item # Model 410N7MC, Model 410N7MC Process Oxygen Analyzer, in Explosion-proof cabinet, Class 1 Div 1 Group BCD



The Nova 410 Series represent Nova's best value in continuous process Oxygen analysis. This analyzer utilizes a long-life electrochemical 'fuel cell' sensor for the detection of O2 in process gases. The sensor produces a small millivoltage output which is directly proportional to the Oxygen detected. The output is then amplified and displayed on a digital readout meter and can also be sent to an optional recorder output. The 410 can be supplied with sampling pump or pressure regulator to suit most applications.

Cabinet: Wall Mounted NEMA7 UL/CSA Explosion-proof with Non-intrusive Magnetic Calibration, Class 1 Div 1 Group BCD. Availability depends on application and environment. Sample conditioning options not available with this cabinet style.





Specifications  · Applications  · Features  · Options  · Calibration

Specifications

Method of Detection

Customer replaceable electrochemical O2 sensor. Expected life 3 to 5 years depending on typical measurement concentrations.

Ranges Available

Any range from 0-2.0% to 0-100.0% O2

Accuracy & Repeatability

Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale

Drift

Less than 2% of full scale per month

Ambient Temperature Range

32 to 122°F (0 to 50°C)

Linearity

Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale

Size & Weight

EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D)

Output Options

4-20 mA into 500 ohms non-isolated standard
Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs

Alarms

High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating
Low flow alarm

Analyzer Type

O2 Analyzer

Primary Gas(es) to be Measured

Oxygen (O2)

Hazardous Area Rating Options

Class 1, Div 1, Grp A-D Hazardous Area Rating


Applications
      For continuous analysis of oxygen (O2) in process gas streams. Also recommended for other applications such as O2 enrichment of combustion air, flue gas analysis, oxygen deficiency analysis, and many more.




Features
  • Best value continuous O2 analyzer
  • Bright digital readout
  • Long-life electrochemical O2 sensor
  • Fast response (T90 - 6 to 8 seconds)
  • Touch-screen or magnetic wand calibration control, depending on cabinet configuration
  • Easy-to-maintain modular layout
  • Built-in sample pump or pressure regulator
  • 4-20mA output




Options
  • Hi/Low gas, low flow alarms available
  • Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
  • Sensors mounted in temperature controlled heated box to improve accuracy




Calibration
  • Nitrogen for zero
  • Air or analyzed calibration gas for span




 MORE IMAGES 

·  Application process


 DOWNLOADS 

410 FlexiNova O2 Analyzer21
(PDF, 575KB)

FlexiNova Ex-Proof Platform Analyzer21
(PDF, 499KB)