Method of Detection |
N/A Customer replaceable electrochemical NO sensor |
Ranges Available |
N/A 0-2000 PPM NOx (as NO) |
Resolution |
N/A 1 PPM NO |
Accuracy & Repeatability |
N/A Better than 2% full scale |
Drift |
N/A Within 1% full scale per 8 hours of continuous operation |
Response Time (T-90) |
N/A 20-30 seconds for 90% of step change |
Ambient Temperature Range |
N/A 12 to 49 ºC55 to 120 ºF |
Linearity |
N/A ±2% of full scale |
Size & Weight |
N/A K Style: 14"L x 10 1/2" W x 6" H @ 12 lbs ( 36 x 27 x 15 cm @ 5.5 kg) WP Style: 11 1/2" L x 8" W x 7 1/4" H @ 8 lbs (29 x 20 x 18 cm @ 3.6 kg) |
Power |
N/A AC/DC Operation, 115 VAC, 60 Hz for recharging (220 VAC 50Hz available) |
Output Options |
N/A 4-20 mA or 0-1 VDC |
Alarms |
N/A NOx alarms with LED |
Unique Applications |
N/A
Note: For proper operation, the sample gas should also contain a small amount of oxygen. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs. |
Primary Gas to be Measured |
N/A Oxides of Nitrogen (NOx) |
Industry Type |
N/A Flue Gas or Emissions |
N/A
|
N/A
|