Print Back
Process PPM Level Carbon Monoxide analyzer from the Nova FLEXI-NOVA Series Platform, designed for light-duty continuous gas measurements.
Continuous Process Gas Analyzers
Unit of Measure
Items Continuous Process Gas Analyzers
Model 480N4
Model 480N4 PPM Carbon Monoxide Analyzer, Wall Mounted NEMA4 (IP65) Enclosure Rating
N4N7 cabinets
Model 480N4N7
Model 480N4N7, Carbon Monoxide Analyzer, ex-proof sensor housing with NEMA4 wall-mount control cabinet (two separate enclosures)
N7MC generic
Model 480N7MC
Model 480N7MC, Carbon Monoxide Analyzer, Wall Mounted NEMA7 UL/CSA Ex-proof with Non-intrusive Magnetic Calibration, Class 1 Div 1 Group BCD
RMN7 21
Model 480RMN7
Model 480RMN7, Carbon Monoxide Analyzer, ex-proof sensor housing with rack mount control cabinet (two separate enclosures)
Purged cabinets
Model 480 D1 or D2
Model 480 D1 or D2, Carbon Monoxide Analyzer - 480 Series; NEMA 4 cabinet with Hazardous Area Purge System
Method of Detection N/A Customer replaceable electrochemical 'fuel cell' CO sensor N/A Customer replaceable electrochemical 'fuel cell' CO sensor. Expected life 1 to 3 years depending on typical measurement concentrations. N/A Customer replaceable electrochemical 'fuel cell' CO sensor. Expected life 1 to 3 years depending on typical measurement concentrations. N/A Customer replaceable electrochemical 'fuel cell' CO sensor. Expected life 1 to 3 years depending on typical measurement concentrations. N/A Customer replaceable electrochemical 'fuel cell' CO sensor. Expected life 1 to 3 years depending on typical measurement concentrations.
Ranges Available N/A Any range from 0-100 PPM to 0-10000 PPM C0
Accuracy & Repeatability N/A ±1% of full scale based on 0-2,000 PPM range N/A ±1% of full scale based on 0-2,000 PPM range N/A ±1% of full scale based on 0-2,000 PPM range N/A ±1% of full scale based on 0-2,000 PPM range N/A ±1% of full scale
Drift N/A Less than 1% of full scale per month or 10 PPM, whichever is greater N/A Less than 1% of full scale per month or 10 PPM, whichever is greater N/A Less than 1% of full scale per month or 10 PPM, whichever is greater N/A Less than 1% of full scale per month or 10 PPM, whichever is greater N/A Less than 1% of full scale per month
Response Time (T-90) N/A Usually 6-8 seconds to 90% step change N/A 6-8 seconds to 90% step change N/A 6-8 seconds to 90% step change N/A 6-8 seconds to 90% step change N/A 6-8 seconds to 90% step change
Ambient Temperature Range N/A 40 to 104°F (4 to 40°C). Optional: up to 130°F (55°C) with cabinet cooler N/A 40 to 104°F (4 to 40°C) N/A 40 to 104°F (4 to 40°C) N/A 40 to 104°F (4 to 40°C) N/A 40 to 104°F (4 to 40°C). Optional: up to 130°F (55°C) with cabinet cooler
Linearity N/A Usually ±1.5% of full scale, depending on range and application
Size & Weight N/A NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D) N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D) N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) RACK MOUNT: 22H x 48W x 47D cm (8.7"H x 19"W x 16"D) N/A NEMA 4: 61H x 51W x 31D cm (24"H x 20"W x 12"D) (not including purge panel & vent)
Power N/A 115 VAC, 60 Hz 220 VAC, 50 Hz
Alarms N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm
Options N/A Cabinet coolers can be fitted to most models Isolated analog, RS232, RS485, serial outputs available Light duty sample conditioning systems available for removal of moisture and/or dust N/A Hi/Low gas, Low Flow, and diagnostic alarms available Isolated 4-20 mA, RS232, RS485 serial outputs N/A Hi/Low gas, Low Flow alarms available Isolated 4-20mA, RS232, RS485, serial outputs N/A Hi/Low gas, Low Flow alarms available Isolated 4-20 mA, RS232, RS485, serial outputs N/A Cabinet coolers can be fitted to most models Cabinet purge system available for use in hazardous areas Hi/Low gas, Low Flow alarms available High temperature filters and probes Isolated 4-20mA, RS232, RS485, serial outputs Light duty sample conditioning systems available for removal of moisture and/or dust
Analyzer Type N/A CO Analyzer (ppm)
Hazardous Area Rating Options N/A General Purpose, Non-Hazardous N/A Control Cabinet - General Purpose, Non-Hazardous Detector Cabinet - NEMA7 Class 1 Division 1 Groups BCD N/A Class 1, Div 1, Grp A-D Hazardous Area Rating N/A Control Cabinet - General Purpose, Non-Hazardous Detector Cabinet - NEMA7 Class 1 Division 1 Groups BCD N/A Class 1 Div 2 or Class 1 Div 1 Hazardous Area Ratings available