In operation, the magneto-dynamic cell makes use of the principle that oxygen is drawn into a magnetic field, thereby increasing the force exerted on a diamagnetic body suspended in the field. The amount of current necessary to restore the position of the body is directly proportional to the O2 level of the sample gas. Very few other gases respond this way, making this method of detection very specific to oxygen.
Cabinets: Ex-proof sensor housing for Class 1 Div 1 Grp BCD areas, with rack mount control cabinet for non-hazardous areas (two separate enclosures)
Specifications
Method of Detection |
N/A Magneto-dynamic paramagnetic O2 cell |
Ranges Available |
N/A With back-pressure control: 1.00, 2.00, 5.00, 25.00, 100.0 % O2 Without back-pressure control: 1.0, 2.0, 5.0 % O2 |
Accuracy & Repeatability |
N/A ±1% of full scale |
Drift |
N/A ±0.2% of full scale per month |
Ambient Temperature Range |
N/A 40 to 104°F (4 to 40°C) |
Linearity |
N/A ±0.5% of full scale or 0.05%, whichever is greater |
Size & Weight |
N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) RACK MOUNT: 22H x 48W x 47D cm (8.7"H x 19"W x 16"D) |
Output Options |
N/A 4-20 mA into 500 ohms non-isolated standard |
Alarms |
N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating |
Analyzer Type |
N/A O2 Analyzer |
Hazardous Area Rating Options |
N/A Control Cabinet - General Purpose, Non-Hazardous Detector Cabinet - NEMA7 Class 1 Division 1 Groups BCD |
Applications
N/A
|
Features
N/A
|