Print Back
The Nova 412 Series O2 analyzers utilize a magneto-dynamic paramagnetic measuring cell and powerful selenium cobalt magnet assembly with on-board pre-amplification to provide the highest standard of O2 analysis. Accuracy is further augmented by built-in barometric pressure compensation and detector temperature control. For high purity applications (Model 412HP), a high resolution digital readout is added.

In operation, the magneto-dynamic cell makes use of the principle that oxygen is drawn into a magnetic field, thereby increasing the force exerted on a diamagnetic body suspended in the field. The amount of current necessary to restore the position of the body is directly proportional to the O2 level of the sample gas. Very few other gases respond this way, making this method of detection very specific to oxygen.

Cabinets: Ex-proof sensor housing for Class 1 Div 1 Grp BCD areas, with rack mount control cabinet for non-hazardous areas (two separate enclosures)
Unit of Measure

Specifications

Method of Detection

N/A Magneto-dynamic paramagnetic O2 cell

Ranges Available

N/A With back-pressure control: 1.00, 2.00, 5.00, 25.00, 100.0 % O2 Without back-pressure control: 1.0, 2.0, 5.0 % O2

Accuracy & Repeatability

N/A ±1% of full scale

Drift

N/A ±0.2% of full scale per month

Ambient Temperature Range

N/A 40 to 104°F (4 to 40°C)

Linearity

N/A ±0.5% of full scale or 0.05%, whichever is greater

Size & Weight

N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) RACK MOUNT: 22H x 48W x 47D cm (8.7"H x 19"W x 16"D)

Output Options

N/A 4-20 mA into 500 ohms non-isolated standard

Alarms

N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating

Analyzer Type

N/A O2 Analyzer

Hazardous Area Rating Options

N/A Control Cabinet - General Purpose, Non-Hazardous Detector Cabinet - NEMA7 Class 1 Division 1 Groups BCD

Applications

N/A

      For continuous analysis of oxygen (O2) in process gas streams that contain gases that interfere with other methods of O2 measurement. For high accuracy or high purity applications.

Features

N/A

  • Highest accuracy O2 analyzer, very stable
  • Fast response (T90 - 4 to 5 seconds) and bright digital readout
  • Linear over entire 0-100% range
  • Microprocessor control
  • Non-consumable O2 detector, no periodic replacement required
  • Built in sample pump or pressure regulator, and 4-20mA output

Options

N/A

  • Hi/Low gas, low flow alarms available
  • Isolated analog, RS485 outputs available
  • High purity version available for 98-100% O2 (412HP)

Calibration

N/A

  • Nitrogen for zero
  • Air or analyzed calibration gas for span

Unique Applications

N/A

    All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.