Method of Detection |
N/A Customer replaceable 'fuel cell' oxygen sensor |
Ranges Available |
N/A Any range between 0-100 PPM and 0-10000 PPM |
Resolution |
N/A 1 PPM |
Accuracy & Repeatability |
N/A ±1% of full scale |
Drift |
N/A Less than 1% of full scale per week |
Response Time (T-90) |
N/A In some cases, ppmO2 sensor may require an initial purge-down from air to low PPM O2 which may last 1-2 hours. After purge, T90 response O2 is less than 20 seconds. |
Ambient Temperature Range |
N/A 40 to 104°F (4 to 40°C) |
Linearity |
N/A ±1% of full scale |
Size & Weight |
N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) RACK MOUNT: 22H x 48W x 47D cm (8.7"H x 19"W x 16"D) |
Power |
N/A 115 VAC, 60 Hz 220 VAC, 50 Hz |
Output Options |
N/A 4-20mA into 500 ohms non-isolated standard Isolated 4-20mA, RS485 outputs available |
Alarms |
N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm |
Analyzer Type |
N/A O2 Analyzer |
Hazardous Area Rating Options |
N/A Control Cabinet - General Purpose, Non-Hazardous Detector Cabinet - NEMA7 Class 1 Division 1 Groups BCD |
N/A Light-duty analyzers for applications or environments that are less intensive. For continuous analysis of PPM levels of oxygen (O2) in clean / dry gas streams. Applications include blanketing mixtures, glove boxes, electronics, research, hydrogen/nitrogen mixtures, and many others. |
N/A
|
N/A
The O2 sensor in the Model 415 is not normally affected by gas stream composition. However, continuous or frequent exposure to high levels of acid gases such as carbon dioxide (CO2) and sulfur dioxide (SO2) can cause decreased cell life. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova at 1-800-295-3771. In many cases, we are able to build an analyzer specific to your needs. |