The Nova 317 Series Portable Hydrocarbon Analyzer is designed for the measurement of propane or gasoline
vapor (using a correction factor) from a closed vessel.

The built-in sampling pump draws in a sample through an external hose where it then passes through a filter, flow
meter and infrared detector for measurement. The measurement results are then displayed on an easy-to-read
digital meter with switchable backlight.

A built-in rechargeable battery has enough power for about 5 hours of continuous operation and the analyzer can
be used while it is recharging. A red LED indicates when to recharge the analyzer and a green LED indicates that
it is receiving recharging power. A recharger is included.
Unit of Measure

Specifications

Method of Detection

N/A NDIR infrared detector (no moving parts)

Ranges Available

N/A 0-100% Propane (other ranges available)

Resolution

N/A 0.1% Propane

Accuracy & Repeatability

N/A ±2% of full scale

Drift

N/A Less than 2% of full scale per week

Response Time (T-90)

N/A 10-15 seconds

Ambient Temperature Range

N/A 0 to 50 ºC32 to 120 ºF

Linearity

N/A ±2% of full scale

Size & Weight

N/A 14" L x 6" H x 10 1/2" D @ 8 lbs (35.5 x 15.2 x 26.6 cm @ 3.6 kg)

Power

N/A AC/DC Operation, 115 VAC / 60 Hz for recharging (220VAC / 50Hz also available)

Output Options

N/A 4-20mA

Alarms

N/A High HC Alarm (optional)

Hazardous Area Rating Options

N/A General Purpose, Non-Hazardous

Applications

N/A

      For analysis of propane or gasoline vapors from a closed vessel or from a closed transfer line, such as with gas station vapor recovery applications.

Features

N/A

  • Rugged design that is easy to operate
  • Fast warm-up and response
  • Solid state infrared detector
  • Calibrates on 100% propane; yields gasoline vapor measurement using a correction factor
  • Digital readout meter with backlight
  • Modular layout that is easy to maintain
  • Rechargeable battery operation
  • Built-in sample pump, filter and flow meter
  • Suitcase style cabinet weatherproof when closed

Options

N/A

  • Recorder output - 4-20mA
  • Suitcase (K) style cabinet available
  • On-board data-logging to USB stick
  • High HC alarm with display indication

Calibration

N/A

  • Air for zero.
  • 100% propane for span.

Unique Applications

N/A This instrument is not explosion-proof or 'intrinsically safe' design. User should observe all locally established precautionary measures when sampling potentially flammable or explosive spaces.