The Nova 410 Series represent Nova's best value in continuous process O2 analysis. This analyzer utilizes a long-life electrochemical 'fuel cell' sensor for the detection of Oxygen in process gases. The sensor produces a small millivoltage output which is directly proportional to the Oxygen detected. The output is then amplified and displayed on a digital readout meter and can also be sent to an optional recorder output. The 410 can be supplied with sampling pump or pressure regulator to suit most applications.

Standard Cabinet: NEMA 4 Weather-proof steel cabinet with window.
Optional: 19" rack mount cabinet also available in some applications.

Specifications

Method of Detection

N/A Customer replaceable electrochemical 'fuel cell' O2 sensor. Expected life 3 to 5 years depending on typical measurement concentrations.

Ranges Available

N/A Any range from 0-2.0% to 0-100.0% O2

Accuracy & Repeatability

N/A Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale

Drift

N/A Less than 2% of full scale per month

Ambient Temperature Range

N/A 32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package

Linearity

N/A Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale

Size & Weight

N/A NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D) RACK MOUNT: 22H x 48W x 47D cm (8.7"H x 19"W x 16"D)

Output Options

N/A 4-20mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS485

Alarms

N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm

Analyzer Type

N/A O2 Analyzer

Primary Gas(es) to be Measured

N/A Oxygen (O2)

Hazardous Area Rating Options

N/A General Purpose, Non-Hazardous

Applications

N/A For continuous analysis of oxygen (O2) in process gas streams. Also recommended for other applications such as O2 enrichment of combustion air, flue gas analysis, oxygen deficiency analysis, and many more.

Features

N/A

  • Best value continuous O2 analyzer
  • Bright digital readout
  • Long life electrochemical O2 sensor
  • Fast response (T90 - 6 to 8 seconds)
  • Microprocessor control for push button calibration
  • Easy-to-maintain modular layout
  • Built-in sample pump or pressure regulator
  • 4-20mA output

Options

N/A

  • Hi/Low gas, and low flow alarms available
  • Isolated 4-20mA, RS485 outputs available
  • Light-duty sample conditioning systems available
  • Cabinet cooler available

Calibration

N/A

  • Nitrogen for zero
  • Air or analyzed calibration gas for span