The Nova FLEXI-NOVA Series Platform has been designed for continuous gas measurements in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series.

Depending on application, some light-duty sample conditioning features may be available. For clean process measurements, percent and ppm-level SO2 gas. All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows connection & data-logging to personal computers via a choice of USB / RS-232 / RS-485.


Cabinets: Wall Mounted NEMA7 UL/CSA Class 1 Division 1 Groups BCD Explosion-proof sensor cabinet with 19" rack mount control cabinet (two separate cabinets)
Unit of Measure

Specifications

Method of Detection

N/A Low range PPM SO2 by electrochemical sensor Microprocessor based single cell NDIR

Ranges Available

N/A 460 - 0-1, 0-5, 0-10% SO2 460L - 0-0-200, 0-1000 PPM SO2 (check availability with Nova)

Resolution

N/A 0.1% SO2 on % versions, 10 PPM SO2 on PPM versions

Accuracy & Repeatability

N/A 1% of full scale

Drift

N/A Less than 2% of full scale per month

Response Time (T-90)

N/A 15-30 seconds to T-90

Ambient Temperature Range

N/A 40 to 122°F (5 to 50°C)

Linearity

N/A ±1% of full scale

Size & Weight

N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D)

Power

N/A 115 VAC, 60 Hz 220 VAC, 50 Hz

Output Options

N/A 4-20mA into 500 ohms non-isolated standard Isolated 4-20mA, RS232, RS485, serial outputs

Alarms

N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm

Analyzer Type

N/A SO2 Analyzer

Hazardous Area Rating Options

N/A Control Cabinet - General Purpose, Non-Hazardous Detector Cabinet - NEMA7 Class 1 Division 1 Groups BCD

Applications

N/A For continuous analysis of sulfur dioxide (SO2) in process gas streams using an infrared detector. Percent and PPM levels available as Models 460 & 460L respectively.

Features

N/A

  • Percent (460) or PPM (460L) SO2 by infrared detector
  • Low range PPM (461) SO2 by electrochemical sensor
  • NDIR detector is sensitive, reliable, and non-depleting
  • Electrochemical sensor is easy to replace when needed
  • Touch-screen display for gas readings
  • Easy-to-maintain modular layout
  • Sensor is temperature controlled for maximum stability
  • Built-in sample pump, flowmeter, and filters
  • 4-20mA non-isolated recorder output

Options

N/A

  • Hi/Low gas and low flow alarms available
  • Serial output & data-logging available

Calibration

N/A

  • Easily zero checked on ambient air
  • On bottled SO2 in N2 gas for Span
  • On bottled N2 or Air for Zero

Unique Applications

N/A Nova 460 / 461 analyzers are suitable for clean dry process gases only. Some gases may interfere with SO2 measurement; complete information on sample gas composition is required. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova at 1-800-295-3771. In many cases, we are able to build an analyzer specific to your needs.