Method of Detection |
N/A Temperature-controlled TC cell. |
Ranges Available |
N/A 0-2.0% to 0-100.0% and truncated ranges such as 85.0-100.0% Consult Nova regarding feasibility of measuring your specific gas mixture. |
Accuracy & Repeatability |
N/A ±1% of full scale |
Drift |
N/A Less than 1% of full scale per month |
Response Time (T-90) |
N/A 10-15 seconds to 90% step change |
Ambient Temperature Range |
N/A 32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package |
Linearity |
N/A ±1% of full scale |
Size & Weight |
N/A Dimensions will vary depending on enclosure style and options required |
Power |
N/A 115 VAC, 60 Hz 220 VAC, 50 Hz |
Output Options |
N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs |
Alarms |
N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm |
Analyzer Type |
N/A Binary Gas Analyzer |
Hazardous Area Rating Options |
N/A General Purpose, Non-Hazardous |
N/A
|
N/A
|
N/A
|