The Nova FLEXI-NOVA Series Platform has been designed for continuous gas measurements in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series.

For clean process measurements of ppm-level CO gas using an electro-chemical sensor. All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows connection & data-logging to personal computers via a choice of USB / RS-232 / RS-485.



Cabinet: NEMA 4 cabinet with Hazardous Area Purge System; available for Class 1 Div1 or Class1 Div2 hazardous areas
Unit of Measure

Specifications

Method of Detection

N/A Customer replaceable electrochemical 'fuel cell' CO sensor. Expected life 1 to 3 years depending on typical measurement concentrations.

Ranges Available

N/A Any range from 0-100 PPM to 0-10000 PPM C0

Resolution

N/A 1 PPM CO up to 2,000 PPM; 10 PPM for greater than 2,000 PPM

Accuracy & Repeatability

N/A ±1% of full scale

Drift

N/A Less than 1% of full scale per month

Response Time (T-90)

N/A 6-8 seconds to 90% step change

Ambient Temperature Range

N/A 40 to 104°F (4 to 40°C). Optional: up to 130°F (55°C) with cabinet cooler

Linearity

N/A Usually ±1.5% of full scale, depending on range and application

Size & Weight

N/A NEMA 4: 61H x 51W x 31D cm (24"H x 20"W x 12"D) (not including purge panel & vent)

Power

N/A 115 VAC, 60 Hz 220 VAC, 50 Hz

Output Options

N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20mA, RS232, RS485, serial outputs

Alarms

N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm

Analyzer Type

N/A CO Analyzer (ppm)

Hazardous Area Rating Options

N/A Class 1 Div 2 or Class 1 Div 1 Hazardous Area Ratings available

Applications

N/A

    For continuous analysis of up to 0-10,000 PPM carbon monoxide (CO) in process gas streams using an electrochemical detector.

Features

N/A

  • Detector is temperature-controlled or compensated for improved analytical stability and fast response
  • PPM range CO measurement by electrochemical sensor
  • Touch-screen display for gas readings
  • Built-in sample pump or pressure regulator
  • Non-isolated 4-20mA analog recorder output

Options

N/A

  • Hi/Low gas and low flow alarms available
  • Serial output & data-logger package available
  • Light-duty pre-filter & condensate removal
  • Up to 6 other gases measured (depending on application)
  • Cabinet cooling available
  • Hazardous area purge systems; available for Class 1 Div1 or Div2 hazardous areas

Calibration

N/A

  • Nitrogen for zero.
  • Analyzed ppmCO calibration gas for span.

Unique Applications

N/A The electrochemical CO sensor is unsuitable with certain background gases such as hydrogen. Also, this cell requires that a small amount of O2 be present in the sample, or that it be metered into the sample. Consult Nova for each application.