Items |
Model 470N7MC Model 470N7MC, Methane Analyzer, Wall Mounted NEMA7 UL/CSA Explosion-proof with Non-intrusive Magnetic Calibration, Class 1 Div 1 Group BCD |
Model 470RMN7 Model 470RMN7, Methane Analyzer, ex-proof sensor housing with rack mount control cabinet (two separate enclosures) |
Model 420 D1 or D2 Model 420 D1 or D2, Methane Analyzer, with cabinet purge systems for hazardous areas |
|||
Method of Detection | N/A Single cell NDIR with no moving parts | |||||
Ranges Available | N/A 470 - 0-1, 0-5, 0-10, 0-20, 0-30, 0-50, 0-100% HCs as CH4 | N/A 0-1, 0-5, 0-10, 0-20, 0-30, 0-50, 0-100% HCs as CH4 | N/A 470 - 0-1, 0-5, 0-10, 0-20, 0-30, 0-50, 0-100% HCs as CH4 | |||
Resolution | N/A 0.1% CH4 on % versions | |||||
Accuracy & Repeatability | N/A 1% of full scale | |||||
Drift | N/A Less than 2% of full scale per month | |||||
Response Time (T-90) | N/A Less than 15 seconds | |||||
Ambient Temperature Range | N/A 40 to 104°F (4 to 40°C) | N/A 40 to 104°F (4 to 40°C) | N/A 40 to 104°F (4 to 40°C). Optional: up to 130°F (55°C) with cabinet cooler | |||
Linearity | N/A ±1% of full scale | |||||
Size & Weight | N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) | N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) RACK MOUNT: 22H x 48W x 47D cm (8.7"H x 19"W x 16"D) | N/A NEMA 4: 61H x 51W x 31D cm (24"H x 20"W x 12"D) (not including purge panel & vent) | |||
Power | N/A 115 VAC, 60 Hz 220 VAC, 50 Hz | |||||
Output Options | N/A 4-20mA into 500 ohms non-isolated standard Isolated 4-20mA, RS232, RS485, serial outputs | N/A 4-20mA into 500 ohms non-isolated standard Isolated 4-20mA, RS232, RS485 serial outputs | N/A 4-20mA into 500 ohms non-isolated standard Isolated 4-20mA, RS232, RS485, serial outputs | |||
Alarms | N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm | |||||
Options | N/A Hi/Low gas, Low Flow alarms available Isolated 4-20 mA, RS232, RS485, serial outputs | N/A Hi/Low gas, Low Flow alarms available Isolated 4-20mA, RS232, RS485 serial outputs | N/A Cabinet coolers can be fitted to most models Cabinet purge system available for use in hazardous areas Hi/Low gas, Low Flow alarms available High temperature filters and probes Isolated 4-20mA, RS232, RS485, serial outputs Light duty sample conditioning systems available for removal of moisture and/or dust | |||
Analyzer Type | N/A Methane Analyzer | N/A Methane Analyzer (HC's as CH4) | N/A Methane Analyzer | |||
Hazardous Area Rating Options | N/A Class 1, Div 1, Grp A-D Hazardous Area Rating | N/A Control Cabinet - General Purpose, Non-Hazardous Detector Cabinet - NEMA7 Class 1 Division 1 Groups BCD | N/A Class 1 Div 2 or Class 1 Div 1 Hazardous Area Ratings available | |||
|