Print Back
The Nova 313 Series Portable Flue Gas Analyzer for NO, NO2, or NOx utilizes reliable, stable NO and/or NO2 sensors which respond quickly to the NO or NO2 present in flue gases or engine exhaust. NO is the major component (90-95%) of the NOx found in flue gases or engine exhaust (except diesel).

In operation, a built in sample pump draws in the gas sample through the S.S. probe, 12 ft sample hose, condensate removal filter, secondary filter and flowmeter, then on to both sensors. The output of
more
Unit of Measure

Specifications

Method of Detection

Customer replaceable electrochemical nitric oxide and nitrogen dioxide sensors Expected life 2-3 year each

Ranges Available

0-2000 PPM NO; 0-800 PPM NO2; 0-2000 PPM Nox; (0-5000 PPM available) Ranges switch selectable

Resolution

1 PPM on 0-2000, 10 PPM on 0-5000

Accuracy & Repeatability

Better than 2% full scale

Drift

Within 1% of full scale per 8 hours of continuous operation

Response Time (T-90)

20-30 seconds

Ambient Temperature Range

12 to 49 ºC55 to 120 ºF

Linearity

±2% of full scale

Size & Weight

approx. 26 x 18 x 13 cm @ 3.4 kg (10" x 7" x 5" @ 7.5lbs)

Power

12V battery operation. 115VAC 60Hz for recharging (220VAC 50Hz available)

Output Options

4-20mA

Alarms

NO, NO2 or NOx alarms with LED

Hazardous Area Rating Options

General Purpose/Non-hazardous

Applications

N/A

        Analysis of oxides of nitrogen (NOx) such as nitric oxide (NO) and/or nitrogen dioxide (NO
2
      ). For boiler, furnace, or engine exhaust monitoring and analysis.

Features

N/A

  • NOx can be read as NO, NO2, or as total
  • Rugged design that is easy to operate
  • Disposable, long life electrochemical NO and NO2 sensors
  • Digital meter readout with backlight
  • Modular layout that is easy to maintain
  • Rechargeable battery operation
  • Built in sample pump, filter, and flowmeter
  • Active condensate removal
  • Rapid reading recovery after NOx 'overdose'
  • Suitcase cabinet (K) weather-proof when closed
  • Stainless steel probe with sample hose

Options

N/A

  • Recorder output - 4-20mA
  • Stack temperature readout (313T)
  • Special sampling probes
  • Sample pre-cooler
  • Suitcase (K) style cabinet available
  • On-board data-logging to USB stick

Calibration

N/A

  • Air for zero.
  • Analyzed calibration gas mixtures of PPM NO and PPM NO2 in nitrogen for span.

Unique Applications

N/A

    All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.