The Nova FLEXI-NOVA Series Platform has been designed for continuous gas measurements in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series. Depending on application, some light-duty sample conditioning features may be available.
For clean process measurements, percent-level O2 gas. This sensor utilizes a paramagnetic measuring cell and powerful selenium cobaltmore magnet assembly to make use of the principle that oxygen is drawn into a magnetic field, thereby offering a method of detection that is very specific to oxygen. Sample dew point should be at least 5°C (9°F) below lowest sample temperature.
All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows connection & data-logging to personal computers via a choice of USB / RS-232 / RS-485.
Cabinet: NEMA 4 cabinet with Hazardous Area Purge System; available for Class 1 Div1 & Div2 hazardous areas less
For clean process measurements, percent-level O2 gas. This sensor utilizes a paramagnetic measuring cell and powerful selenium cobaltmore magnet assembly to make use of the principle that oxygen is drawn into a magnetic field, thereby offering a method of detection that is very specific to oxygen. Sample dew point should be at least 5°C (9°F) below lowest sample temperature.
All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows connection & data-logging to personal computers via a choice of USB / RS-232 / RS-485.
Cabinet: NEMA 4 cabinet with Hazardous Area Purge System; available for Class 1 Div1 & Div2 hazardous areas less
Unit of Measure
Specifications
Method of Detection |
N/A Magneto-dynamic paramagnetic O2 cell |
Ranges Available |
N/A With back-pressure control: 1.00, 2.00, 5.00, 25.00, 100.0 % O2 Without back-pressure control: 1.0, 2.0, 5.0 % O2 |
Accuracy & Repeatability |
N/A ±0.5% of full scale or 0.05%, whichever is greater |
Drift |
N/A ±0.2% of full scale per month |
Ambient Temperature Range |
N/A 40 to 104°F (4 to 40°C). Optional: up to 130°F (55°C) with cabinet cooler |
Linearity |
N/A ±0.5% of full scale or 0.05%, whichever is greater |
Size & Weight |
N/A Dimensions will vary depending on purge system and options required NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D) |
Output Options |
N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs |
Alarms |
N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating |
Analyzer Type |
N/A O2 Analyzer |
Hazardous Area Rating Options |
N/A Class 1 Div 2 or Class 1 Div 1 Hazardous Area Ratings available |
Applications
N/A
|
Features
N/A
|
Options
N/A
|
View Our
Product Catalog