Cabinet: Wall Mounted NEMA7 UL/CSA Explosion-proof with Non-intrusive Magnetic Calibration, Class 1 Div 1 Group BCD.
Availability depends on application and environment. Sample conditioning options not available with this cabinet style. more
Unit of Measure

Specifications

Method of Detection

Customer replaceable 'fuel cell' oxygen sensor

Ranges Available

Any range between 0-100 PPM and 0-10000 PPM

Accuracy & Repeatability

±1% of full scale

Drift

Less than 1% of full scale per week

Ambient Temperature Range

40 to 104°F (4 to 40°C)

Linearity

±1% of full scale (minimum 0.5 PPM)

Size & Weight

EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D)

Output Options

4-20mA into 500 ohms non-isolated standard Isolated 4-20mA, RS485

Alarms

High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm

Analyzer Type

O2 Analyzer

Hazardous Area Rating Options

Class 1, Div 1, Grp A-D Hazardous Area Rating

Applications

N/A Light-duty analyzers for applications or environments that are less intensive. For continuous analysis of PPM levels of oxygen (O2) in clean / dry gas streams. Applications include blanketing mixtures, glove boxes, electronics, research, hydrogen/nitrogen mixtures, and many others.

Features

N/A

  • Bright digital readout in PPM O2
  • Ranges available from 0-100 PPM to 0-10000 PPM O2
  • Unaffected by hydrocarbons in the sample
  • Easy-to-maintain, modular layout
  • Built in sample pump or pressure regulator, and 4-20mA outputs
  • Manual or solenoid valves trap nitrogen (N2) in cell to avoid depletion when analyzer is powered off
  • Touch-screen display for gas readings and for start-up programming at installation time
  • Magnetic switches allow subsequent calibration and operation control (on standalone N7MC version)

Options

N/A

  • Hi/Low gas and low flow alarms available
  • Serial output available

Calibration

N/A

  • Nitrogen for zero
  • Air or analyzed calibration gas for span

Unique Applications

N/A The O2 sensor in the Model 415 is not normally affected by gas stream composition. However, continuous or frequent exposure to high levels of acid gases such as carbon dioxide (CO2) and sulfur dioxide (SO2) can cause decreased cell life. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova at 1-800-295-3771. In many cases, we are able to build an analyzer specific to your needs.