The Nova 510 Series oxygen deficiency or excess alarm monitors are designed as an O2 'watchdog' against accidental displacement of ambient air in a confined space by inert gases, or by a buildup of too much O2 from an O2 leak in the room. The monitors all utilize a long life electrochemical O2 sensor. Actual O2 concentration is shown on a digital meter on the front panel.

Optional dual level alarm with local indication, relay contacts, and audible alert available. Alarm set points are pre-set at Nova but are field adjustable if desired. Flashing strobe light and upgraded cabinets also available.



Models:
  • 510D - Diffusion sensor, no pump and no moving parts; excellent for O2 monitoring in immediate local environment
  • 510P - Pumped sample version; pump pulls air sample from wider area or from another isolated environment, such as from a closed room
Unit of Measure

Specifications

Method of Detection

N/A Customer replaceable electrochemical oxygen sensor

Ranges Available

N/A 0-30% O2

Resolution

N/A 0.1% O2

Accuracy & Repeatability

N/A ± 2% of full scale

Drift

N/A Less than 1% of full scale per month

Response Time (T-90)

N/A 6-8 seconds to 90% step change

Ambient Temperature Range

N/A 40° to 104°F (5° to 40°C)

Linearity

N/A ±2% of full scale

Size & Weight

N/A 20" H x 16" W x 10" D @ 20lbs. (51 x 41 x 26 cm @ 10kg)

Power

N/A 115 VAC, 60 Hz 220 VAC, 50 Hz

Output Options

N/A 4-20mA non-isolated

Alarms

N/A Optional High and/or Low O2 alarm with SPDT contacts with 10A 115VAC rating

Analyzer Type

N/A Ambient O2 analyzer

Area Rating

N/A General Purpose Non-Hazardous only

Applications

N/A To monitor confined spaces, labs, control rooms, tunnels, oxygen plants, etc. for oxygen (O2) deficiency or excess.

Features

N/A

  • Best value high accuracy O2 monitor
  • Bright digital readout
  • 4 to 5 year life electrochemical O2 sensor
  • Available with pumped or diffusion-style sampling
  • Easy to use and calibrate
  • Fast response (T90 - 6 to 8 seconds)
  • NEMA 12 dust tight enclosure
  • 4-20mA recorder output

Options

N/A

  • Two levels of O2 alarm with local indication, relay contacts, and audible alarm with 'Acknowledge' button
  • Low flow alarm on pumped sample models
  • NEMA 4 weather-resistant enclosure
  • Strobe light for alarm indication
  • 510D - Diffusion sensor, no pump and no moving parts; excellent for O2 monitoring in immediate local environment
  • 510P - Pumped sample version; pump pulls air sample from wider area or from another isolated environment, such as from a closed room

Calibration

N/A

  • Although calibration of zero should not be necessary, it can be checked on nitrogen
  • Analyzed gas or outside ambient air with 20.9% O2 for span

Unique Applications

N/A All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova at 1-800-295-3771. In many cases, we are able to build an analyzer specific to your needs.