Sample Gas is drawn into the analyzer from the sampling probe, then pumped through a condensate removal system. Any condensate formed collects and is purged as it reaches a limit. Next, the dried sample gas is filtered again and then flows through a PTFE liquid block, calibration gas valves, pressure regulator, flow meter then on to the gas detector.
The 7200-1 analyzer features an electrochemical sensor, paramagnetic sensor or zirconia sensor for detecting % levels of O2, a digital readout and an output for the O2 channel. The 7200-1 analyzer is multi-sample point capable (consult Nova).
Specifications
Gases Measured |
N/A Oxygen (O2) |
Method of Detection |
N/A Electrochemical sensor, paramagnetic sensor or zirconia sensor (depending on desired accuracy and sample composition) |
Ranges Available |
N/A 0-5 to 0-25% O2 |
Resolution |
N/A 0.1% on % ranges, 1 PPM on PPM ranges |
Accuracy & Repeatability |
N/A ± 1-2% of full scale |
Drift |
N/A Less than 2% of full scale per month |
Response Time (T-90) |
N/A 20-30 seconds to 90% of step change |
Ambient Temperature Range |
N/A 32 to 122°F (0 to 50°C) @ 5-95% RH non-condensing. Lower temperatures (-5°F, -20°C) with Cold Weather Package. Sample dew point should be at least 9°F (5°C) below lowest sample temperature. |
Linearity |
N/A ±1% of full scale |
Size & Weight |
N/A Dimensions and weight will vary with configurations selected |
Power |
N/A 115 VAC, 60 Hz 220 VAC, 50 Hz |
Alarms |
N/A High or low alarm contacts available for all gases and low flow |
Code Numbers for Gases Measured |
N/A O2 by electrochemical sensor, paramagnetic sensor or zirconia sensor (depending on desired accuracy and sample composition) |
Sampling System |
N/A Std SS probe with unheated pre-filter (Std. probe 3/8" O.D. x 18") (9.5 x 460 mm) |
Display Type |
N/A D1 - LED display (s) - Manual potentiometer calibration D2 - LED display (s) - Manual push-button calibration D3 - Microprocessor controller with single graphic display. Allows auto sequencing, RS-485 output, 'CAL NOW' and one-step D4 - Same as D3 but with full auto-calibration |
Output Options |
N/A OP1 - 4-20 mA non-isolated OP2 - 4-20 mA isolated OP3 - 0-5 V OP4 - 0-10 V OP5 - RS-485 OP6 - Modbus |
Analyzer Type |
N/A Flue Gas Analyzer |
Industry Type |
N/A Flue Gas or Emissions |
Applications
N/A
|
Features
N/A
|
Options
N/A
|