The Nova FLEXI-NOVA Series Platform has been designed for continuous gas measurements in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series. Depending on application, some light-duty sample conditioning features may be available.
For clean process measurements, percent-level O2 gas. This sensor utilizes a paramagnetic measuring cell and powerful selenium cobalt magnet assembly to make use of the principle that oxygen is drawn into a magnetic field, thereby offering a method of detection that is very specific to oxygen. Sample dew point should be at least 5°C (9°F) below lowest sample temperature.
All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls.
Cabinet: NEMA 7 ex-proof cabinet, Class 1 Div 1 Group BCD
For clean process measurements, percent-level O2 gas. This sensor utilizes a paramagnetic measuring cell and powerful selenium cobalt magnet assembly to make use of the principle that oxygen is drawn into a magnetic field, thereby offering a method of detection that is very specific to oxygen. Sample dew point should be at least 5°C (9°F) below lowest sample temperature.
All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls.
Cabinet: NEMA 7 ex-proof cabinet, Class 1 Div 1 Group BCD
Unit of Measure
Specifications
Method of Detection |
N/A Magneto-dynamic paramagnetic O2 cell |
Ranges Available |
N/A With back-pressure control: 1.00, 2.00, 5.00, 25.00, 100.0 % O2 Without back-pressure control: 1.0, 2.0, 5.0 % O2 |
Resolution |
N/A 0.1% or 0.01% O2, depending on back-pressure control |
Accuracy & Repeatability |
N/A ±1% of full scale |
Drift |
N/A ±0.2% of full scale per month |
Response Time (T-90) |
N/A 4 - 5 seconds at a sample flow of 1 LPM |
Ambient Temperature Range |
N/A 32 to 122°F (0 to 50°C) @ 5-95% RH non-condensing. Lower temperatures (-5°F, -20°C) with Cold Weather Package. Sample dew point should be at least 9°F (5°C) below lowest sample temperature. |
Linearity |
N/A ±0.5% of full scale or 0.05% absolute, whichever is greater, depending on back-pressure control |
Size & Weight |
N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) |
Power |
N/A 115 VAC, 60 Hz 220 VAC, 50 Hz |
Output Options |
N/A 4-20mA into 500 ohms non-isolated standard |
Alarms |
N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating |
Analyzer Type |
N/A O2 Analyzer |
Hazardous Area Rating Options |
N/A Class 1, Div 1, Grp A-D Hazardous Area Rating |
Applications