For clean process measurements, percent-level H2 gas. This analyzer utilizes a temperature-controlled, thermalconductivity (TC) cell for the detection of H2 in process gas. The cell produces a small millivoltage output which is directly proportional to the H2 detected. The analyzer can be supplied with a pump or pressure regulator depending on customer requirement.
All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows local USB connection & data-logging to personal computers.
Cabinet: Wall Mounted NEMA7 UL/CSA Explosion-proof with Non-intrusive Magnetic Calibration, Class 1 Div 1 Group BCD
Specifications
Method of Detection |
N/A Temperature-controlled TC cell. Expected life in excess of 10 years. Cell cannot be burned out due to loss of sample flow |
Ranges Available |
N/A Any range between 0-2% and 0-100% H2 |
Resolution |
N/A 0.1% H2 (0.01% H2 available) |
Accuracy & Repeatability |
N/A ±1% of full scale |
Drift |
N/A Less than 1% of full scale per month |
Response Time (T-90) |
N/A 10-15 seconds to 90% step change |
Ambient Temperature Range |
N/A 40 to 104°F (4 to 40°C) |
Linearity |
N/A ±1% of full scale |
Size & Weight |
N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) |
Power |
N/A 115 VAC, 60 Hz 220 VAC, 50 Hz |
Output Options |
N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20mA, RS232, RS485, serial outputs |
Alarms |
N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating |
Analyzer Type |
N/A H2 Analyzer |
Hazardous Area Rating Options |
N/A Class 1, Div 1, Grp A-D Hazardous Area Rating |
Applications
N/A Light-duty analyzers for applications or environments that are less intensive. For continuous analysis of up to 100% hydrogen (H2) in clean process gas streams. |