New Call-to-action
The Nova FLEXI-NOVA Series Platform has been designed for continous gas measurments in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series.

Depending on application, some light-duty sample conditioning features may be available. For clean process measurements, of certain binary gas mixtures such as H2 or He in Air. This analyzer utilizes a temperature-controlled, thermalconductivity (TC) cell for the required measurement. The cell produces a small millivoltage output which is directly proportional to the gas detected. The analyzer can be supplied with a pump or pressure regulator depending on customer requirement.

All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows USB connection & data-logging to personal computers.

Cabinet: NEMA 4 cabinet with Hazardous Area Purge System; available for Class 1 Div1 & Div2 hazardous areas

Unit of Measure

Specifications

Method of Detection

N/A Temperature-controlled TC cell

Ranges Available

N/A 0-2.0% to 0-100.0% and truncated ranges such as 85.0-100.0% Consult Nova regarding feasibility of measuring your specific gas mixture.

Accuracy & Repeatability

N/A Usually ±1% of full scale

Drift

N/A Usually less than 1% of full scale per month

Response Time (T-90)

N/A Usually 10-15 seconds to 90% step change

Ambient Temperature Range

N/A 40 to 104°F (4 to 40°C). Optional: up to 130°F (55°C) with cabinet cooler.

Linearity

N/A Usually ±1% of full scale

Size & Weight

N/A NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D) (not including purge panel & vent)

Power

N/A 115 VAC, 60 Hz 220 VAC, 50 Hz

Output Options

N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, USB serial output

Alarms

N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm

Analyzer Type

N/A Binary Gas Analyzer

Hazardous Area Rating Options

N/A Class 1 Div 2 or Class 1 Div 1 Hazardous Area Ratings available

Applications

N/A Light-duty analyzers for applications or environments that are less intensive. For continuous analysis of a choice of many binary gas mixtures comprised of H2, CO2, O2, CH4, SO2, Ar, He, NH3, SF6, Air, and others. Applications include heat treating atmospheres, welding gas mixtures, blanketing gas mixtures, ammonia synthesis, gas purge monitoring, magnesium blanketing by SF6, and others.

Features

N/A

  • Bright digital readout
  • Lower cost than infrared, GC, or mass spectrometer
  • High stability, microprocessor-controlled, long-life thermal conductivity (TC) cell
  • Easy-to-maintain modular layout
  • Built-in sample pump or pressure regulator
  • 4-20mA recorder output
  • Sensors temperature-controlled for maximum stability

Options

N/A

  • Hi/Low gas and low flow alarms available
  • Serial output & data-logger package available
  • Light-duty pre-filter & condensate removal
  • Cabinet cooling available for NEMA4 cabinet
  • Class 1 Div 2 or Class 1 Div 1 Hazardous Area Ratings available using cabinet positive pressure purge system

Calibration

N/A

  • In most cases, on bottled gas for Span and Zero

Unique Applications

N/A

    All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.

View Our
Product Catalog
Have a question?
Contact Us
Ready to get started?
Request Quote