The Nova 257 Series Continuous Dew Point Analyzers have been designed primarily for accurate dew point measurement of exothermic or endothermic heat treating atmospheres.

The dielectric polymer sensor provides stable and accurate dew point measurements and is not affected by most gas compositions.

A built in sample pump draws in samples that are of insufficient pressure from the process. The sample flow is shown on a small rotameter with valve. The analyzer then reads the dew point continuously.
Unit of Measure

Specifications

Method of Detection

N/A Dielectric polymer sensor that changes capacitance with moisture variation

Ranges Available

N/A -40°F to +104°F (-40°C to +40°C) Dew Point

Resolution

N/A ±1°F (1°C) Dew Point

Accuracy & Repeatability

N/A ±1°F (±1°C) Dew Point

Drift

N/A Less than 1% of full scale per month

Response Time (T-90)

N/A Less than 60 seconds after initial purge down

Ambient Temperature Range

N/A -18 to 49 ºC0 to 120 ºF

Linearity

N/A Better than 1°F (1°C) Dew Point

Size & Weight

N/A 16" W x 14" H x 8" D @ 10 lbs (41 x 36 x 20 cm @ 4.5 kg)

Power

N/A 115 VAC, 60 Hz for recharging 220 VAC, 50 Hz available

Output Options

N/A 4-20 mA or 0-1 VDC

Alarms

N/A High or low dew point alarms with LED (optional)

Hazardous Area Rating Options

N/A General Purpose, Non-Hazardous

Applications

N/A

    For continuous dew point analysis at atmospheric pressure of air, heat treat furnace gases, compressed air lines and other applications with oils or contaminants present.

Features

N/A

  • Standard range of -40 to +68°F (-40 to +20°C) dew point. Other ranges and units of measure available on request
  • Rugged design that is easy to operate
  • Use on any non-corrosive gas
  • Long-life dielectric polymer sensor
  • May be calibrated in the field
  • Bright digital readout
  • Built in sample pump, filter, and flow meter
  • NEMA 12 enclosure is standard
  • Sensor easily replaced if contaminated
  • Factory calibrated to NIST traceable standards

Options

N/A

  • Recorder outputs of 0-1V or 4-20mA
  • Calibration kit
  • Also available in a NEMA 12 or 19" rack mount cabinets
  • Carbon potential calibration using input from customer supplied Type 'K' thermocouple

Calibration

N/A

  • Calibration is easily performed in the field using the optional calibration kit (Part no. NDPCK-02)
  • Alternatively, the analyzer may be returned to Nova for a certified calibration

Unique Applications

N/A

    The 257 Series should not be used to measure dew points when corrosive gases are present. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.