New Call-to-action
The Nova FLEXI-NOVA Series Platform has been designed for continuous gas measurements in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series.

Depending on application, some light-duty sample conditioning features may be available. For clean process measurements of percent-level CO gas by infrared detector.

All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows connection & data-logging to personal computers via a choice of USB / RS-232 / RS-485.



Cabinets: Wall Mounted NEMA7 UL/CSA Class1 Div1 Explosion-proof sensor cabinet, with NEMA4 wall-mount control cabinet (two separate cabinets)

Unit of Measure

Specifications

Method of Detection

N/A Single cell NDIR with no moving parts

Ranges Available

N/A High ranges - any from 0-40 to 0-100% CO Standard ranges - 0-1.00, 2.00, 5.00, 10, 30% CO

Resolution

N/A 0.1% on most % ranges

Accuracy & Repeatability

N/A ±1-3% of full scale, depending on measurement range and application

Drift

N/A Usually 2% of full scale per month or less, depending on range and application

Response Time (T-90)

N/A Usually 20-30 seconds to 90% step change, depending on range and application

Ambient Temperature Range

N/A 40 to 104°F (4 to 40°C)

Linearity

N/A Usually ±1.5% of full scale, depending on range and application

Size & Weight

N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D)

Power

N/A 115 VAC, 60 Hz 220 VAC, 50 Hz

Output Options

N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, serial outputs

Alarms

N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm

Analyzer Type

N/A CO analyzer (IR)

Hazardous Area Rating Options

N/A Control Cabinet - General Purpose, Non-Hazardous Detector Cabinet - NEMA7 Class 1 Division 1 Groups BCD

Applications

N/A Light-duty analyzers for applications or environments that are less intensive. For continuous analysis of percent levels of carbon monoxide (CO) in process gas streams using an infrared detector.

Features

N/A

  • Detector is temperature-controlled or compensated for improved analytical stability
  • Standard ranges, any between 0-30% CO
  • High ranges available (up to 100%)
  • Touch-screen display for gas readings
  • Built-in sample pump or pressure regulator
  • Non-isolated 4-20mA analog recorder output

Options

N/A

  • Hi/Low gas and low flow alarms available
  • Serial output & data-logging available

Calibration

N/A

  • Easily Zero-checked on ambient air
  • Span using bottled CO gas in N2

Unique Applications

N/A The FlexiNova 481 Series analyzers are designed for detection of CO in non-corrosive gas streams. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova at 1-800-295-3771. In many cases, we are able to build an analyzer specific to your needs.

View Our
Product Catalog
Have a question?
Contact Us
Ready to get started?
Request Quote