New Call-to-action
The Nova 910 Series analyzers have been designed for monitoring recovered and processed gases from landfill sites. This equipment continuously removes moisture from the wet sample gas without the need for cooling water or compressed air.

The 910 Series comes complete with pre-filter, sample pump, flow meter, continuous moisture removal system, long-life O2 sensor, infrared CO2/CH4 detector, low flow sensor, alarm relay contacts, and 4-20mA outputs for each gas measured.

Sensors / detectors are mounted in an internal dedicated enclosure which is heated and temperature controlled to reduce drift due to ambient temperature swings. The internal sensor enclosure is removable to allow quick field change of sensors. The infrared detector can be cleaned in the field if necessary by the on-site technician. Environmental management packages are available to allow operation of analyzer outdoors in temperatures from -30°C (-22°F) up to 55°C (131°F).


All calibration functions are controlled through the display rather than through manual potentiometers. The operator only has to input the calibration frequency and leave calibration gas cylinder connected to the analyzer - the rest is automatic. Diagnostics and troubleshooting information are also included in the display interface. This assists operators to quickly determine the cause of an alarm and get the analyzer back on line.


Models
  • 910C - O2
  • 911C - CH4
  • 912C - O2 and CH4
  • 913C - O2, CH4, and CO2
  • 914C - CH4 and CO2



Unit of Measure

Specifications

Method of Detection

N/A O2 by electrochemical sensor, CH4 & CO2 by infrared detector

Ranges Available

N/A 0-25.0% O2; any range between 0-50.0% to 0-100.0% CH4, CO2

Resolution

N/A 0.1% on CO2, O2, & CH4

Accuracy & Repeatability

N/A ±1% of full scale

Drift

N/A Less than 2% of full scale per month

Response Time (T-90)

N/A Less than 30 seconds not including sample line delay

Ambient Temperature Range

N/A 40°F to 104°F (4°C to 40°C), Optional: -22°F to 158°F (-30°C to 70°C) with Outdoor Packages

Linearity

N/A ±1% of full scale for each gas measured

Size & Weight

N/A Approx. 24" H x 24" W x 10" D @ 90 lbs (61 x 61 x 25 cm @ 41 kg); physical data may vary depending on options required.

Power

N/A 115 VAC 60 Hz (220 VAC 50 Hz available)

Output Options

N/A Isolated 4-20 mA standard RS232, RS485, MODBUS®, Ethernet outputs optional

Alarms

N/A Optional: High O2, high CO2, low CH4, low flow; SPDT relays with 10A contacts

Analyzer Type

N/A Landfill Gas Analyzer

Primary Gas(es) to be Measured

N/A O2 / CO2 / CH4

Hazardous Area Rating Options

N/A Class 1 Div 1 or Class 2 Div 2, using purge system General Purpose, Non-Hazardous

Applications

N/A For continuous analysis of oxygen (O2), methane (CH4), and/or carbon dioxide (CO2) in recovered gases from landfill sites.

Features

N/A

  • Automatic analyzer calibration with color touch screen LCD display
  • 'Cal-Now' feature initiates full calibration with a single swipe of magnetic wand
  • Temperature-controlled sensors for maximum stability
  • Long-life electrochemical O2 sensor
  • Infrared detector for reliable measurement of CH4 & CO2
  • Stainless steel hard tubing and corrosion-resistant components for internal sample path
  • Cabinet positive-pressure system to minimize the corrosive effects of hydrogen sulfide (H2S) on cabinet electronics & sensors
  • Built-in sample pump, pre-filter, filters, and flow meter
  • Automatic condensate removal
  • Back pressure regulator minimizes vacuum effects if the sample is vented back to blower
  • Rugged NEMA4 (IP65) enclosure
  • Isolated 4-20mA output for each gas
  • Low flow alarm & relay

Options

N/A

  • Gas High/Low alarms with relay contacts
  • RS232, RS485, MODBUS®, and Ethernet outputs
  • Outdoor weather packages for operation from -30°C (-22°F) up to 55°C (131°F)
  • In-cabinet LEL monitor to interrupt analyzer power in the event of an internal gas leak

Calibration

N/A

  • Ambient air for O2 span and CO2 & CH4 zero
  • Analyzed calibration gas of CO2 and CH4 in nitrogen for CO2 & CH4 span and O2 zero

View Our
Product Catalog
Have a question?
Contact Us
Ready to get started?