New Call-to-action
The Nova FLEXI-NOVA Series Platform has been designed for continuous gas measurements in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series.

For clean process measurements, percent-level H2 gas. This analyzer utilizes a temperature-controlled, thermalconductivity (TC) cell for the detection of H2 in process gas. The cell produces a small millivoltage output which is directly proportional to the H2 detected. The analyzer can be supplied with a pump or pressure regulator depending on customer requirement.

All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows local USB connection & data-logging to personal computers.


Cabinet: Wall Mounted NEMA7 UL/CSA Explosion-proof with Non-intrusive Magnetic Calibration, Class 1 Div 1 Group BCD
Unit of Measure

Specifications

Method of Detection

N/A Temperature-controlled TC cell. Expected life in excess of 10 years. Cell cannot be burned out due to loss of sample flow

Ranges Available

N/A Any range between 0-2% and 0-100% H2

Resolution

N/A 0.1% H2 (0.01% H2 available)

Accuracy & Repeatability

N/A ±1% of full scale

Drift

N/A Less than 1% of full scale per month

Response Time (T-90)

N/A 10-15 seconds to 90% step change

Ambient Temperature Range

N/A 40 to 104°F (4 to 40°C)

Linearity

N/A ±1% of full scale

Size & Weight

N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D)

Power

N/A 115 VAC, 60 Hz 220 VAC, 50 Hz

Output Options

N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20mA, RS232, RS485, serial outputs

Alarms

N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating

Analyzer Type

N/A H2 Analyzer

Hazardous Area Rating Options

N/A Class 1, Div 1, Grp A-D Hazardous Area Rating

Applications

N/A Light-duty analyzers for applications or environments that are less intensive. For continuous analysis of up to 100% hydrogen (H2) in clean process gas streams.

Features

N/A

  • Best value continuous H2 analyzer
  • Local digital readout, 4-20mA recorder output
  • Long life thermal conductivity (TC) detector with high accuracy and repeatability, cannot be burned out from loss of flow
  • Easy-to-maintain, modular layout, fast response
  • Built-in sample pump or pressure regulator
  • Sensor temperature-controlled for maximum stability
  • Ex-proof Class1 Div1 cabinet, with Non-intrusive Magnetic Calibration

Options

N/A

  • Hi/Low gas, low flow, and diagnostic alarms available
  • Isolated analog, RS232, RS485 serial outputs available

Calibration

N/A

  • Nitrogen or air for zero
  • Span using a known H2 in nitrogen calibration gas.

Unique Applications

N/A

    All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.