New Call-to-action
The Nova 430 Series Process Hydrogen analyzer utilizes a temperature-controlled TC cell for the detection of H2 in process gas. This cell produces a small millivoltage output which is directly proportional to the H2 detected. This output is then amplified and displayed on a digital readout meter and is also available as an output recorder. The analyzer can be supplied with a pump or pressure regulator depending on customer requirement.

Cabinet: NEMA 4 Weather-proof steel cabinet with window
Unit of Measure

Specifications

Method of Detection

N/A Temperature-controlled TC cell. Expected life in excess of 10 years. Cell cannot be burned out due to loss of sample flow

Ranges Available

N/A Any range between 0-2% and 0-100% H2

Resolution

N/A 0.1% H2 (0.01% H2 available)

Accuracy & Repeatability

N/A ±1% of full scale

Drift

N/A Less than 1% of full scale per month

Response Time (T-90)

N/A 10-15 seconds to 90% step change

Ambient Temperature Range

N/A 32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package. Dew point should be at least 9°F (5°C) below lowest sample temperature

Linearity

N/A ±1% of full scale

Size & Weight

N/A Dimensions will vary depending on enclosure style and options required

Power

N/A 115 VAC, 60 Hz 220 VAC, 50 Hz

Output Options

N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs

Alarms

N/A High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating

Analyzer Type

N/A H2 Analyzer

Hazardous Area Rating Options

N/A General Purpose, Non-Hazardous

Applications

N/A

    For continuous analysis of up to 100% hydrogen (H2) in process gas streams. Also recommended for other applications such as H2 cooled generators, electrolysis units, refinery gas, heat treating atmospheres, blast furnace gas, and many more.

    The Nova 430 Series Process Hydrogen analyzer utilizes a temperature-controlled TC cell for the detection of H2 in process gas. This cell produces a small millivoltage output which is directly proportional to the H2 detected. This output is then amplified and displayed on a digital readout meter and is also available as an output recorder. The analyzer can be supplied with a pump or pressure regulator depending on customer requirement.

Features

N/A

  • Best value continuous H2 analyzer
  • Bright digital readout, 4-20 mA outputs
  • Long life thermal conductivity (TC) detector with high accuracy and repeatability, cannot be burned out from loss of flow
  • Fast response (T90 - 10 to 15 seconds)
  • Easy-to-maintain modular layout
  • Built in sample pump or pressure regulator
  • Sensor mounted in temperature controlled heated box to improve accuracy

Options

N/A

  • Hi/Low gas, low flow, and diagnostic alarms available
  • Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
  • Sample conditioning systems available for removal of moisture, acidic gas, and dust
  • Cabinet purge system available for use in hazardous areas
  • One-step 'cal-now' button with touch screen LCD display
  • Full automatic calibration with touch screen LCD display
  • Cold weather package for operation to -5°F (-20°C)
  • Cabinet coolers can be fitted to most models
  • High temperature filters and probes

Calibration

N/A

  • Nitrogen or air for zero
  • Span using a known H2 in nitrogen calibration gas.

Unique Applications

N/A

    All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.

View Our
Product Catalog
Have a question?
Contact Us
Ready to get started?