Method of Detection |
N/A Customer replaceable electrochemical oxygen and CO sensors Solid state infra-red detector for CO2 |
Ranges Available |
N/A 0-1800°F (0-1000°C) stack temperature 0-20.0% CO2 0-2000 PPM and 0-4.00% CO - Switch selectable 0-25.0% Oxygen |
Resolution |
N/A 0.1% O2, 1 PPM on CO and CO2 |
Accuracy & Repeatability |
N/A 2% of full scale for O2 and CO2; ±10 PPM CO |
Drift |
N/A <2% full scale per 8 hours of continuous operation |
Response Time (T-90) |
N/A 10-15 seconds for O2; 20-30 seconds for CO and CO2 |
Ambient Temperature Range |
N/A 0 to 40 ºC32 to 105 ºF |
Linearity |
N/A ±1.0% of full scale for each gas measured |
Size & Weight |
N/A 16" L x 4 1/2" H x 8" D @ 13 lbs (40.6 x 20.3 x 10.8 cm @ 5.6 kg) |
Power |
N/A AC/DC Operation, 115 VAC 60 Hz for recharging (Other voltages available.) |
Output Options |
N/A 4-20 mA or 0-1 VDC |
Unique Applications |
N/A The 375 Series should not be used for detecting these gases in ambient atmospheres for personnel safety purposes. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs. |
N/A
|
N/A
|