In operation, the TC cell detects the rate at which the sample gas conducts heat away from a Resistance Temperature Device (RTD) with reference to a similar RTD surrounded by air (or other reference gas). A sample gas containing the gas to be measured will cause the temperature, and its resistance, to change with respect to the reference RTD. This will cause an output from the Wheatstone bridge measuring circuit which is calibrated and displayed on a digital meter.
Method of Detection |
N/A CH4 in N2, air, or CO2; SF6 in CO2, etc. CO2 in N2, H2, air, Ar, O2, He, etc O2 in Ar, CO2, He, H2, etc RTD-based TC cell can be used to measure almost any binary gas mixture such as H2 in N2, O2, air, SO2, Ar, He, NH3, etc |
Ranges Available |
N/A 0-2% to 0-100% and suppressed ranges such as 95-100% |
Resolution |
N/A 0.1% of gas to be measured |
Accuracy & Repeatability |
N/A ±1% of full scale |
Drift |
N/A Less than 1% of full scale per month |
Response Time (T-90) |
N/A 10-15 seconds to 90% step change |
Ambient Temperature Range |
N/A 32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package |
Linearity |
N/A ±1% of full scale |
Size & Weight |
N/A Dimensions will vary depending on enclosure style and options required |
Power |
N/A 115 VAC, 60 Hz 220 VAC, 50 Hz |
Output Options |
N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs |
Alarms |
N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm |
Unique Applications |
N/A
|
Primary Gas to be Measured |
N/A Ammonia (NH3) Argon Carbon Dioxide (CO2) Helium (He) Hydrogen (H2) Methane (CH4) Nitrogen (N2) Oxygen (O2) Sulfur Dioxide (SO2) Sulfur Hexaflouride (SF6) |
Industry Type |
N/A Online Process Analysis |
Mounting Options |
N/A Panel mounted Rack mounted Wall mounted |
Models and Enclosure Options |
N/A NEMA12 (IP52) NEMA4 (IP65) NEMA4X (IP65) NEMA7 NEMA7 UL/CSA |
N/A
|
N/A
|
N/A
|
N/A
|
Product Catalog
Contact Us