In operation, a built in sample pump draws in the flue gas sample through the stainless steel probe, 12 ft. (4 m) sample hose, condensate removal filter, secondary filter, flow meter, then on to the O2, CO and combustibles sensors. The detected gases are displayed on LCD digital meters which have a switchable backlight for use in dark areas. A built in air makeup system ensures that the combustibles detector will always have sufficient O2 for proper operation regardless of sample O2 content.
A rechargeable battery provides enough power for about 6 hours of continuous operation and the analyzer can be used while it is being recharged. A red LED tells when to recharge and a green LED verifies that it is receiving recharging power. The recharger is included.
The Nova 352T indicates stack temperature for doing fuel efficiency calculations. The temperature sensor is built into the sampling probe. Efficiency charts for each fuel are provided
Method of Detection |
N/A Catalytic oxidation detector for combustibles; expected life is 3-4 years for each Customer replaceable electrochemical O2 and CO sensors |
Ranges Available |
N/A 0-1800°F or 0-1000 °C stach temperature (Model 352T) 0-2000PPM and 4.00% CO (switch selectable) 0-25.0% O2 0-5.0% or 0-10.0% combustibles |
Resolution |
N/A 0.1% O2, CO and combustibles |
Accuracy & Repeatability |
N/A ±1% full scale, based on 20.9% O2; ± 2% of full scale combustibles |
Drift |
N/A <2% full scale per 8 hours of continuous operation |
Response Time (T-90) |
N/A 5-8 seconds for O2; 20-30 seconds for combustibles |
Ambient Temperature Range |
N/A 0 to 40.5 ºC32 to 105 ºF |
Linearity |
N/A ±1% full scale O2 and CO; ±2% of full scale combustibles |
Size & Weight |
N/A Suitcase (K): 14" L x 10 1/2" W x 6" H @ 12 lbs (36 x 27 x 15 cm @ 5.5 kg) Weather Proof (WP) Case: 10" L x 7 1/2" H x 8" W @ 8 lbs (25 x 19 x 20 cm @ 3.6 kg) |
Power |
N/A AC/DC Operation, 115 VAC, 60 Hz for recharging (Other voltages available) |
Output Options |
N/A 4-20 mA or 0-1 VDC |
Unique Applications |
N/A
The 352 Series should not be used for detecting these gases in ambient atmospheres for personnel safety purposes. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs. |
Primary Gas to be Measured |
N/A Flue Gas or Emissions |
Industry Type |
N/A Steel Making |
N/A
|
N/A
|
N/A
|
N/A
|