Continuous Process Oxygen Gas Analyzer - 410 Series
Continuous Oxygen Analyzers for Hazardous Locations (O2)
For continuous analysis of Oxygen (O2) in process gas streams. Also recommended for other applications such as O2 enrichment of combustion air, Flue Gas Analysis, Oxygen deficiency analysis, and many more.
Unit of Measure
Items |
/Asset/N7MC-generic.jpg Model 410N7MC Model 410N7MC Process Oxygen Analyzer, in Explosion-proof cabinet, Class 1 Div 1 Group BCD |
/Asset/RMN7-21.jpg Model 410RMN7 Model 410RMN7 Process Oxygen Analyzer, ex-proof sensor housing with rack mount control cabinet (two separate enclosures) |
/Asset/N4-+-D1-&-D2-purge21.jpg Model 410 D1 or D2 Model 410 D1 or D2 Oxygen Analyzer with cabinet purge systems for hazardous areas |
|||
Method of Detection | N/A Customer replaceable electrochemical O2 sensor. Expected life 3 to 5 years depending on typical measurement concentrations. | N/A Customer replaceable electrochemical 'fuel cell' O2 sensor. Expected life 3 to 5 years depending on typical measurement concentrations. | N/A Customer replaceable electrochemical 'fuel cell' O2 sensor. Expected life 3 to 5 years depending on typical measurement concentrations. | |||
Ranges Available | N/A Any range from 0-2.0% to 0-100.0% O2 | |||||
Resolution | N/A 0.1% O2 | |||||
Accuracy & Repeatability | N/A Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale | |||||
Drift | N/A Less than 2% of full scale per month | |||||
Response Time (T-90) | N/A 6-8 seconds to 90% step change | |||||
Ambient Temperature Range | N/A 32 to 122°F (0 to 50°C) | N/A 32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package | N/A 32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package | |||
Linearity | N/A Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale | |||||
Size & Weight | N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) | N/A EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D) RACK MOUNT: 22H x 48W x 47D cm (8.7"H x 19"W x 16"D) | N/A Dimensions will vary depending on type of cabinet purge required NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D) | |||
Power | N/A 115 VAC, 60 Hz 220 VAC, 50 Hz | |||||
Output Options | N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs | |||||
Alarms | N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm | |||||
Options | N/A High temperature filters and probes One-step 'cal-now' button with touch screen LCD display | N/A High temperature filters and probes Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available | N/A Cabinet coolers can be fitted to most models Cabinet purge system available for use in hazardous areas Hi/Low gas, Low Flow, and diagnostic alarms available High temperature filters and probes Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available Sample conditioning systems available for removal of moisture and/or dust | |||
Analyzer Type | N/A O2 Analyzer | |||||
Primary Gas(es) to be Measured | N/A Oxygen (O2) | |||||
Hazardous Area Rating Options | N/A Class 1, Div 1, Grp A-D Hazardous Area Rating | N/A Control Cabinet - General Purpose, Non-Hazardous Detector Cabinet - NEMA7 Class 1 Division 1 Groups BCD | N/A Class 1 Div 2 or Class 1 Div 1 Hazardous Area Ratings available | |||
|
View Our
Product Catalog
Product Catalog
Have a question?
Contact Us
Contact Us
Ready to get started?
Request Quote
Request Quote