New Call-to-action
Analysis of Oxides of Nitrogen (NOx) such as Nitric Oxide (NO) and/or Nitrogen Dioxide (NO2). For boiler, furnace, or engine exhaust monitoring and analysis.
Unit of Measure
Items
313-WP-NOx

Portable Flue Gas Analyzer for Oxides Of Nitrogen, Weatherproof (WP) Enclosure

K-1-gas

Portable Flue Gas Analyzer for Oxides Of Nitrogen, Suitcase (K) Enclosure
Method of Detection N/A Customer replaceable electrochemical nitric oxide and nitrogen dioxide sensors Expected life 2-3 year each
Ranges Available N/A 0-2000 PPM NO; 0-800 PPM NO2; 0-2000 PPM Nox; (0-5000 PPM available) Ranges switch selectable N/A 0-2000 PPM NO; 0-800 PPM NO2; 0-2000 PPM Nox; (0-5000 PPM available) Ranges switch selectable
Resolution N/A 1 PPM on 0-2000, 10 PPM on higher ranges N/A 1 PPM on 0-2000, 10 PPM on 0-5000
Accuracy & Repeatability N/A Better than 2% full scale
Drift N/A Within 1% of full scale per 8 hours of continuous operation
Response Time (T-90) N/A 20-30 seconds
Ambient Temperature Range N/A 12 to 49 ºC55 to 120 ºF
Linearity N/A ±2% of full scale
Size & Weight N/A 10" L x 7 1/2" H x 8" D @ 8 lbs (25 x 19 x 20 cm @ 3.6 kg) N/A 14" L x 10 1/2" W x 6" H @ 12 lbs (36 x 27 x 15 cm @ 5.5 kg)
Power N/A AC/DC Operation, 115 VAC 60 Hz for recharging (Other voltages available)
Output Options N/A 4-20 mA or 0-1 VDC
Options N/A AC power only operation Data Logger NO, NO2 or NOx alarms with LED Recorder outputs of 0-1 V or 4-20 mA (can be isolated) Sample pre-cooler Special sampling probes Stack temperature readout (313T) Suitcase (K) style cabinet available
Analyzer Type N/A NOx Analyzer
Alarms N/A NO, NO2 or NOx alarms with LED
Hazardous Area Rating Options N/A General Purpose/Non-hazardous