New Call-to-action
For checking dew points of at atmospheric pressure of air, heat treat furnace gases, compressed air lines and other application with oils or contaminants present.
Unit of Measure
Items
AA Portable Gas Analyzers

Model 207K-KX, Portable Standard Range Dew Point Analyzer

Remote Transmitter Series 202

Model 207RT, Remote Dew Point Transmitter with Flow-Through Chamber
Method of Detection N/A Dielectric polymer sensor that change capacitance with moisture variation
Ranges Available N/A -40°C to +40°C (or -40°F to +104°F) Dew Point N/A -40°C to +40°C Dew Point -40°F to +104°F Dew Point
Resolution N/A 1°F (1°C) Dew Point
Accuracy & Repeatability N/A ±1°F (±1°C) Dew Point
Drift N/A Less than 1% of full scale range per month
Response Time (T-90) N/A Less than 30 seconds
Ambient Temperature Range N/A 4° to 49°C ºC40° to 120°F ºF N/A -18 to 49 ºC0 to 120 ºF
Linearity N/A Better than 1°F (1°C) Dew Point
Size & Weight N/A K Suitcase Style - Approx. 14" L x 6" H x 11" D @ 12 lbs (36 x 15 x 27 cm @ 5.4 kg N/A Approx. 10" L x 6" W x 4" D @ 5 lbs. (26 x 25 x 10 cm @ 2.3 kg)
Power N/A 115 VAC 60 Hz for recharging 220VAC 50Hz available N/A 12 to 24 VDC
Output Options N/A 4-20mA N/A 4-20 mA
Options N/A Calibration kit Carbon potential calculation (202-CP) Output display in relative humidity (202RH) N/A Calibration kit
Alarms N/A High or low dew point alarms with LED (optional)
Hazardous Area Rating Options N/A General Purpose, Non-Hazardous
View Our
Product Catalog
Have a question?
Contact Us
Ready to get started?