Continuous analysis of flue gas for any combination and range of O2, CO, CO2, SO2, NOx (NO or NO + NO2) or combustibles from combustion processes.
Unit of Measure
Items
7200-1Gas

Continuous Flue Gas Analyzer - 7200 Series

7200-1Gas

Continuous Flue Gas Analyzer - 7200 Series

7200-1Gas

Continuous Flue Gas Analyzer - 7200 Series

7200-1Gas

Continuous Flue Gas Analyzer - 7200 Series

7200-1Gas

Continuous Flue Gas Analyzer - 7200 Series
Method of Detection N/A Electrochemical sensor, paramagnetic sensor or zirconia sensor (depending on desired accuracy and sample composition) N/A PPM CO by electrochemical sensor, % CO by NDIR infrared detector N/A Infrared detector N/A PPM NO by electrochemical sensor N/A Catalytic sensor (incl. air make up)
Ranges Available N/A 0-5 to 0-25% O2 N/A 0-100, 0-250, 0-500, 0-2000 by E/C, 0-10,000 PPM CO by infrared detector N/A 0-20% CO2 N/A 0-100 to 0-5000 PPM NOx (as NO) N/A 0-2, 0-5, 0-10% Combustibles
Resolution N/A 0.1% on % ranges, 1 PPM on PPM ranges N/A 0.1% on %, 1 PPM on PPM N/A 0.1% on %, 1 PPM on PPM N/A 1 PPM N/A 0.1%
Accuracy & Repeatability N/A ± 1-2% of full scale
Drift N/A Less than 2% of full scale per month
Response Time (T-90) N/A 20-30 seconds to 90% of step change
Ambient Temperature Range N/A 32 to 122°F (0 to 50°C) @ 5-95% RH non-condensing. Lower temperatures (-5°F, -20°C) with Cold Weather Package. Sample dew point should be at least 9°F (5°C) below lowest sample temperature.
Linearity N/A ±1% of full scale
Size & Weight N/A Dimensions and weight will vary with configurations selected
Power N/A 115 VAC, 60 Hz 220 VAC, 50 Hz
Alarms N/A High or low alarm contacts available for all gases and low flow
Code Numbers for Gases Measured N/A O2 by electrochemical sensor, paramagnetic sensor or zirconia sensor (depending on desired accuracy and sample composition) N/A CO by electrochemical sensor or infrared detector N/A CO2 by infrared detector N/A NOx (as NO or NO + NOPPM NO by electrochemical sensor N/A Combustibles by catalytic sensor (incl. air make up)
Sampling System N/A Std SS probe with unheated pre-filter (Std. probe 3/8" O.D. x 18") (9.5 x 460 mm)
Display Type N/A D1 - LED display (s) - Manual potentiometer calibration D2 - LED display (s) - Manual push-button calibration D3 - Microprocessor controller with single graphic display. Allows auto sequencing, RS-485 output, 'CAL NOW' and one-step D4 - Same as D3 but with full auto-calibration
Output Options N/A OP1 - 4-20 mA non-isolated OP2 - 4-20 mA isolated OP3 - 0-5 V OP4 - 0-10 V OP5 - RS-485 OP6 - Modbus
Analyzer Type N/A Flue Gas Analyzer
Options N/A Cabinet coolers can be fitted to most models Cabinet purge system available for use in hazardous areas Cold weather package for operation to -5°F (-20°C) Full automatic calibration with touch screen LCD display and low flow switch Hi/Low gas, low flow, and diagnostic alarms available High temperature filters and probes Isolated analog, RS-232, RS-485, MODBUS, and Ethernet outputs available One-step 'cal-now' button with touch screen LCD display Sensors mounted in heated box for reduced drift