New Call-to-action
For checking purity of Hydrogen (H2) in H2-cooled generators and synchronous converters. The Model 380 Series will also monitor the safe purging of H2 during shut down or startup.
1 - 3 of 3 | Results Per Page | View | Unit of Measure
1 

Item #

Enclosure and Mounting Details

Ranges Available

Resolution

Drift

Hazardous Area Rating Options

N/A Portable Tri Gas Analyzer for Hydrogen Purging, Suitcase (K) Enclosure N/A Range 1 : 0-100% H2 in Air Range 2 : 0-100% H2 in CO2 Range 3 : 0-100% Air in CO2 N/A 0.1% of gas measured N/A H2 in Air range is ± 0.4% per week H2 in CO2 or Air in CO2 2% F.S. per week maximum drift N/A General Purpose, Non-Hazardous
N/A Portable Tri-Gas Analyzer for H2 / N2 / Air Purging N/A Range 1: 0-100% H2 in Air
Range 2: 0-100% H2 in N2
Range 3: 0-100% Air in N2
N/A 0.1% of gas measured N/A H2 in N2 or Air in N2, ± 2% F.S. per week maximum drift; 0-100% H2 in Air range is ± 0.4% per week maximum N/A General Purpose, Non-Hazardous
N/A Portable Tri Gas Analyzer for Hydrogen Purging with separate O2 channel N/A Range 1: 0-100% H2 in Air
Range 2: 0-100% H2 in CO2
Range 3: 0-100% Air in CO2
Separate Range: 0-25.0 % O2
N/A 0.1% of gas measured N/A H2 in CO2, Air in CO2, & O2 ± 2% F.S. per week maximum drift; 0-100% H2 in Air range is ± 0.4% per week maximum N/A General Purpose, Non-Hazardous
1 - 3 of 3 | Results Per Page | View | Unit of Measure
1 
View Our
Product Catalog
Have a question?
Contact Us
Ready to get started?
Request Quote