All Categories > By Primary Gas to be Measured > Oxygen Analyzers (O2) > Continuous Oxygen Analyzers for Hazardous Locations (O2) > Continuous Process Oxygen Gas Analyzer by Paramagnetic Sensor - 412 Series > Item # Model 410 D1 or D2  


Item # Model 410 D1 or D2, Model 410 D1 or D2, Oxygen Analyzer with cabinet purge systems for hazardous areas


The Nova FLEXI-NOVA Series Platform has been designed for continuous gas measurements in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series. Depending on application, some light-duty sample conditioning features may be available.

For clean process measurements, percent-level O2 gas. This sensor utilizes a paramagnetic measuring cell and powerful selenium cobalt magnet assembly to make use of the principle that oxygen is drawn into a magnetic field, thereby offering a method of detection that is very specific to oxygen. Sample dew point should be at least 5°C (9°F) below lowest sample temperature.

All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows connection & data-logging to personal computers via a choice of USB / RS-232 / RS-485.

Cabinet: NEMA 4 cabinet with Hazardous Area Purge System; available for Class 1 Div1 & Div2 hazardous areas






Specifications  · Applications  · Features  · Options  · Calibration  · Unique Applications

Specifications

Method of Detection

Magneto-dynamic paramagnetic O2 cell

Ranges Available

With back-pressure control: 1.00, 2.00, 5.00, 25.00, 100.0 % O2
Without back-pressure control: 1.0, 2.0, 5.0 % O2

Resolution

±1% of full scale

Accuracy & Repeatability

±0.5% of full scale or 0.05%, whichever is greater

Drift

±0.2% of full scale per month

Response Time (T-90)

4 - 5 seconds at a sample flow of 1 LPM

Ambient Temperature Range

40 to 104°F (4 to 40°C). Optional: up to 130°F (55°C) with cabinet cooler

Linearity

±0.5% of full scale or 0.05%, whichever is greater

Size & Weight

Dimensions will vary depending on purge system and options required
NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D)

Power

115 VAC, 60 Hz
220 VAC, 50 Hz

Output Options

4-20 mA into 500 ohms non-isolated standard
Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs

Alarms

High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating

Analyzer Type

O2 Analyzer

Hazardous Area Rating Options

Class 1 Div 2 or Class 1 Div 1 Hazardous Area Ratings available


Applications
      For continuous analysis of oxygen (O2) in process gas streams that contain gases that interfere with other methods of O2 measurement. For high accuracy or high purity applications.




Features
  • Highest accuracy O2 analyzer, very stable
  • Fast response (T90 - 4 to 5 seconds) and bright digital readout
  • Linear over entire 0-100% range
  • Microprocessor control
  • Non-consumable O2 detector, no periodic replacement required
  • Built in sample pump or pressure regulator, and 4 - 20 mA outputs
  • To improve accuracy, detector can be mounted in temperature controlled heated box and compensated for changes in barometric pressure at the vent




Options
  • Hi/Low gas, low flow alarms available
  • Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
  • Sample conditioning systems available for removal of moisture and dust
  • Cabinet purge system available for use in hazardous areas
  • Cabinet coolers can be fitted to most models
  • High purity version available for 98-100% O2 (412HP)




Calibration
  • Nitrogen for zero
  • Air or analyzed calibration gas for span




Unique Applications
    All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.