Method of Detection
|
|
Customer replaceable electrochemical 'fuel cell' O2 sensor. Expected life 3 to 5 years depending on typical measurement concentrations.
|
|
Ranges Available
|
|
Any range from 0-2.0% to 0-100.0% O2
|
|
Resolution
|
|
0.1% O2
|
|
Accuracy & Repeatability
|
|
Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale
|
|
Drift
|
|
Less than 2% of full scale per month
|
|
Response Time (T-90)
|
|
6-8 seconds to 90% step change
|
|
Ambient Temperature Range
|
|
32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package
|
|
Linearity
|
|
Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale
|
|
Size & Weight
|
|
Dimensions will vary depending on type of cabinet purge required NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D)
|
|
Power
|
|
115 VAC, 60 Hz 220 VAC, 50 Hz
|
|
Output Options
|
|
4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs
|
|
Alarms
|
|
High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm
|
|
Analyzer Type
|
|
O2 Analyzer
|
|
Primary Gas(es) to be Measured
|
|
Oxygen (O2)
|
|
Hazardous Area Rating Options
|
|
Class 1 Div 2 or Class 1 Div 1 Hazardous Area Ratings available
|
|