Canada
270 Sherman Avenue N.
Hamilton, ON L8L 6N5
Phone: (905) 545 2003• Fax: (905) 545 4248
United States
1925 Pine Avenue
Niagara Falls, NY 14301
Phone: (716) 285-0418 • Fax: (716) 282-2937
 
Email: sales@nova-gas.com or websales@nova-gas.com
All Categories > By Primary Gas to be Measured > Hydrogen Analyzers (H2) > Continuous Hydrogen Analyzers (H2) > Continuous Process Hydrogen Gas Analyzer - 430 Series > Item # Model 430RMN7  


Item # Model 430RMN7, Model 430RMN7, Hydrogen Analyzer, ex-proof sensor housing with rack mount control cabinet (two separate enclosures)


The Nova 430 Series Process Hydrogen analyzer utilizes a temperature-controlled TC cell for the detection of H2 in process gas. This cell produces a small millivoltage output which is directly proportional to the H2 detected. This output is then amplified and displayed on a digital readout meter and is also available as an output recorder. The analyzer can be supplied with a pump or pressure regulator depending on customer requirement.

Cabinets: Wall Mounted NEMA7 UL/CSA Explosion-proof sensor cabinet with 19" rack mount control cabinet (two separate cabinets)






Specifications  · Applications  · Features  · Options  · Calibration  · Unique Applications

Specifications

Method of Detection

Temperature-controlled TC cell. Expected life in excess of 10 years. Cell cannot be burned out due to loss of sample flow

Ranges Available

Any range between 0-2% and 0-100% H2

Resolution

0.1% H2 (0.01% H2 available)

Accuracy & Repeatability

±1% of full scale

Drift

Less than 1% of full scale per month

Response Time (T-90)

10-15 seconds to 90% step change

Ambient Temperature Range

32 to 122°F (0 to 50°C)

Linearity

±1% of full scale

Size & Weight

EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D)
RACK MOUNT: 22H x 48W x 47D cm (8.7"H x 19"W x 16"D)

Power

115 VAC, 60 Hz
220 VAC, 50 Hz

Output Options

4-20 mA into 500 ohms non-isolated standard
Isolated 4-20 mA, RS232, RS485, serial outputs

Alarms

High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating

Analyzer Type

H2 Analyzer

Hazardous Area Rating Options

Control Cabinet - General Purpose, Non-Hazardous
Detector Cabinet - NEMA7 Class 1 Division 1 Groups BCD


Applications

Light-duty analyzers for applications or environments that are less intensive. For continuous analysis of up to 100% hydrogen (H2) in clean process gas streams.





Features
  • Best value continuous H2 analyzer
  • Local digital readout, 4-20mA recorder output
  • Long life thermal conductivity (TC) detector with high accuracy and repeatability, cannot be burned out from loss of flow
  • Easy-to-maintain, modular layout, fast response
  • Built-in sample pump or pressure regulator
  • Sensor temperature-controlled for maximum stability




Options
  • Hi/Low gas and low flow alarms available
  • Serial output & data-logger package available
  • Light-duty pre-filter
  • 19" rack mount cabinet control cabinet




Calibration
  • Nitrogen or air for zero
  • Span using a known H2 in nitrogen calibration gas.




Unique Applications
    All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.





Print Back