Item # Model 480N7MC, Model 480N7MC, Carbon Monoxide Analyzer, Wall Mounted NEMA7 UL/CSA Ex-proof with Non-intrusive Magnetic Calibration, Class 1 Div 1 Group BCD
|
|
|
|
The Nova FLEXI-NOVA Series Platform has been designed for continuous gas measurements in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series. For clean process measurements of ppm-level CO gas using an electro-chemical sensor. All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows connection & data-logging to personal computers via a choice of USB / RS-232 / RS-485.
Cabinet: Wall Mounted NEMA7 UL/CSA Explosion-proof with Non-intrusive Magnetic Calibration, Class 1 Div 1 Group BCD
|
|
|
|
|
Method of Detection
|
|
Customer replaceable electrochemical 'fuel cell' CO sensor. Expected life 1 to 3 years depending on typical measurement concentrations.
|
|
Ranges Available
|
|
Any range from 0-100 PPM to 0-10000 PPM C0
|
|
Resolution
|
|
1 PPM CO up to 2,000 PPM; 10 PPM for greater than 2,000 PPM
|
|
Accuracy & Repeatability
|
|
±1% of full scale based on 0-2,000 PPM range
|
|
Drift
|
|
Less than 1% of full scale per month or 10 PPM, whichever is greater
|
|
Response Time (T-90)
|
|
6-8 seconds to 90% step change
|
|
Ambient Temperature Range
|
|
40 to 104°F (4 to 40°C)
|
|
Linearity
|
|
Usually ±1.5% of full scale, depending on range and application
|
|
Size & Weight
|
|
EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D)
|
|
Power
|
|
115 VAC, 60 Hz 220 VAC, 50 Hz
|
|
Output Options
|
|
4-20 mA into 500 ohms non-isolated standard Isolated 4-20mA, RS232, RS485, serial outputs
|
|
Alarms
|
|
High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm
|
|
Analyzer Type
|
|
CO Analyzer (ppm)
|
|
Hazardous Area Rating Options
|
|
Class 1, Div 1, Grp A-D Hazardous Area Rating
|
|
|
|
|
For continuous analysis of up to 0-10,000 PPM carbon monoxide (CO) in process gas streams using an electrochemical detector.
|
|
|
|
- Detector is temperature-controlled or compensated for improved analytical stability and fast response
- PPM range CO measurement by electrochemical sensor
- Touch-screen display for initial set-up
- Built-in sample pump or pressure regulator
- Non-isolated 4-20mA analog recorder output
- Ex-proof Class1 Div1 cabinet, with Non-intrusive Magnetic Calibration
|
|
|
|
- Hi/Low gas and low flow alarms available
- Serial output & data-logger package available
|
|
|
|
- Nitrogen for zero.
- Analyzed ppmCO calibration gas for span.
|
|
|
|
The electrochemical CO sensor is unsuitable with certain background gases such as hydrogen. Also, this cell requires that a small amount of O2 be present in the sample, or that it be metered into the sample. Consult Nova for each application.
|
|
|
|
|
|
|
|
|
|
|
|