Method of Detection |
N/A Customer-replaceable electrochemical ppm oxygen sensor |
Ranges Available |
N/A Any range between 0-100 PPM and 0-10,000 PPM |
Accuracy & Repeatability |
N/A ±1% of full scale |
Drift |
N/A Less than 1% of full scale per week |
Ambient Temperature Range |
N/A 40 to 104°F (4 to 40°C). Optional: up to 130°F (55°C) with cabinet cooler |
Linearity |
N/A ±1% of full scale (minimum 0.5 PPM) |
Size & Weight |
N/A NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D) RACK MOUNT: 22H x 48W x 47D cm (8.7"H x 19"W x 16"D) |
Output Options |
N/A 4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs |
Alarms |
N/A High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm |
Analyzer Type |
N/A O2 Analyzer |
Hazardous Area Rating Options |
N/A General Purpose, Non-Hazardous |
N/A Light-duty analyzers for applications or environments that are less intensive. For continuous analysis of PPM levels of oxygen (O2) in clean / dry gas streams. Applications include blanketing mixtures, glove boxes, electronics, research, hydrogen/nitrogen mixtures, and many others. |
N/A
|
N/A
|
N/A
The O2 sensor in the Model 415 is not normally affected by gas stream composition. However, continuous or frequent exposure to high levels of acid gases such as carbon dioxide (CO2) and sulfur dioxide (SO2) can cause decreased cell life. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova at 1-800-295-3771. In many cases, we are able to build an analyzer specific to your needs. |