Item # Model 415N4N7, Model 415N4N7 Trace Oxygen Analyzer, , ex-proof sensor housing with NEMA4 wall-mount control cabinet (two separate enclosures)
|
|
|
|
The Nova FLEXI-NOVA Series Platform has been designed for continuous gas measurements in applications and environments that are less intensive. Lab work, research, clean processes, pre-treated / pre-cleaned produced gases, and others, may benefit from the FLEXI-NOVA series.
Depending on application, some light-duty sample conditioning features may be available. For clean process measurements, PPM-level O2 gas. This electrochemical type sensor is unaffected by most other gases in the sample including most hydrocarbons and acidic gases such as carbon dioxide (CO2), hydrogen sulfide (H2S), and sulfur dioxide (SO2). Typical O2 sensor life is 1-2 years or longer depending on measurement concentrations and sample gas exposure. The sensor may be easily replaced by the customer.
All sensors / detectors are temperature-controlled or temperature-compensated for maximum analytical stability. Easy calibration using touch-screen controls. Serial Output Package allows connection & data-logging to personal computers via a choice of USB / RS-232 / RS-485.
Cabinets: (two separate cabinets)
Detector Cabinet - Wall Mounted NEMA7 UL/CSA Explosion-proof Class 1 Div1 Grp BCD Control Cabinet - NEMA4 wall-mount non-hazardous areas
|
|
|
|
|
|
|
Light-duty analyzers for applications or environments that are less intensive. For continuous analysis of PPM levels of oxygen (O2) in clean / dry gas streams. Applications include blanketing mixtures, glove boxes, electronics, research, hydrogen/nitrogen mixtures, and many others.
|
|
|
|
- Touch-screen display for gas readings
- Ranges available from 0 to 100 PPM to 0 to 9,999 PPM O2
- Electrochemical O2 sensor is unaffected by hydrocarbons in the sample
- Solenoid valves trap PPM O2 sample to avoid sensor depletion when analyzer is powered off
- Easy-to-maintain, modular layout
- Non-isolated 4-20mA recorder output
|
|
|
|
- Hi/Low gas and low flow alarms available
- Serial output available
- Light-duty pre-filter & condensate removal
- Cabinet cooling available (for control cabinet only)
|
|
|
|
- Nitrogen for zero
- Air or analyzed calibration gas for span
|
|
|
|
The O2 sensor in the Model 415 is not normally affected by gas stream composition. However, continuous or frequent exposure to high levels of acid gases such as carbon dioxide (CO2) and sulfur dioxide (SO2) can cause decreased cell life. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova at 1-800-295-3771. In many cases, we are able to build an analyzer specific to your needs.
|
|
|
|
|
|
|
|
|
|
|
|
|
|