All Categories > By Primary Gas to be Measured > Oxygen Analyzers (O2) > Continuous Oxygen Analyzers (O2) > Continuous Biogas Recovery & Processing Analyzers - 920 Series > Item # 920D2 Series  


Item # 920D2 Series, Continuous Analyzer for Biogas Recovery & Processing - Methane + Oxygen + Carbon Dioxide; NEMA 4 cabinet with Hazardous Area Purge System


The Nova 920 Series analyzers have been designed for monitoring recovered and processed gases from biogas installations. The Model 923 is specifically designed for continuous oxygen (O2), carbon dioxide (CO2), and methane (CH4) measurement. The sample conditioning system continuously removes moisture from the wet sample gas without the need for cooling water, compressed air, or failure-prone thermoelectric elements. The 920 Series comes complete with pre-filter, sample pump, flow meter, continuous moisture removal system, long-life electrochemical O2 sensor, infrared CO2 & HC's detector (as CH4), low flow sensor, alarm relay contacts, and 4-20mA output.

Sensors / detectors are mounted in an internal dedicated enclosure which is temperature-controlled to reduce drift due to ambient temperature swings. The internal sensor enclosure is removable to allow quick field change of sensors. The infrared detector can be cleaned in the field if necessary by the on-site technician.

Cabinet: NEMA 4 cabinet with Hazardous Area Purge System; available for Class 1 Div1 & Div2 hazardous areas

 





 Specifications 

Ranges Available

0-25.0 % O2
Any range between 0-50.0% and 0-100.0% CH4, CO2