All Categories > By Primary Gas to be Measured > Helium Analyzers (He) > Portable Process Helium Gas Analyzer - 334 Series > Item # Model 334K  


Item # Model 334K, Portable Process Analyzer for Helium, Suitcase (K) Enclosure


The Nova 334 Portable He Analyzer has been designed for the detection of He primarily in air or nitrogen.

The thermal conductivity (T/C) cell provides a fast and accurate measurement of He. It has an expected life of over 10 years unless contaminated.

In operation, a built in sample pump draws in the gas sample through the probe, filters, flow meter, and on to the T/C cell. The detected He is displayed on a digital readout meter.

A rechargeable battery provides power for approximately 8 hours of operation and the analyzer can be used while it is being recharged. The recharger is included.






Specifications  · Applications  · Features  · Options  · Calibration  · Unique Applications

Specifications

Method of Detection

Temperature compensated thermal conductivity (T/C) cell cannot be burned out due to loss of flow or changing gases

Ranges Available

0-100% He in air or in N2

Resolution

0.1% of He

Accuracy & Repeatability

±2% of full scale

Drift

±1% of full scale per week

Response Time (T-90)

90% of final reading in 10 seconds

Ambient Temperature Range

12 to 50 ºC
55 to 120
ºF

Linearity

±2% of F.S. on He in Air

Size & Weight

K style - approx. 35.5L x 15.2H x 26.6D cm @ 5.5 kg (14" x 6" x 10½" @ 8 lbs)

Power

115 VAC / 60 Hz for recharging (220 VAC / 50 Hz available)

Output Options

4-20mA

Hazardous Area Rating Options

General Purpose, Non-Hazardous


Applications
For analysis of Helium (He) in a binary gas mixture in process gases such as He in air, He in Nitrogen (N2), He in Carbon Dioxide (CO2), He in Argon (Ar), He in Oxygen (O2), etc.



Features
  • Rugged design that is easy to operate
  • Fast warm up and response
  • Long-life thermal conductivity cell that provides accurate and stable readings
  • Digital meter readout
  • Modular layout that is easy to maintain
  • Rechargeable battery operated
  • Built in sample pump
  • Lightweight, compact and easy to use
  • Suitcase cabinet (K), weatherproof when closed




Options
  • Recorder output - 4-20mA
  • Sample pre-cooler for hot samples
  • Suitcase (K) style cabinet available




Calibration
  • Ambient air for zero
  • 100% He calibration gas for span.




Unique Applications
All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.