Canada
270 Sherman Avenue N.
Hamilton, ON L8L 6N5
Phone: (905) 545 2003• Fax: (905) 545 4248
United States
1925 Pine Avenue
Niagara Falls, NY 14301
Phone: (716) 285-0418 • Fax: (716) 282-2937
 
Email: sales@nova-gas.com or websales@nova-gas.com
All Categories > By Primary Gas to be Measured > Dew Point Analyzers (DP) > Portable Dew Point Analyzers & Testers > Portable Low Range Dewpoint Analyzer - 203K-KX > Item # Model 203K-KX  


Item # Model 203K-KX, Model 203K-KX Portable Low Range Dewpoint Analyzer


The Model 203 Series Portable Dew Point Analyzers have been designed primarily for accurate dew point measurement of dry non-corrosive industrial gases, or dry oil-free air. The aluminum oxide sensing element varies its electrical capacitance in direct proportion to variations in the sample gas dew point.

The 203 Series can have either a built-in pressure regulator for positive pressure samples, or a built-in sample pump for atmospheric pressures, or both. The sample flow is shown on a flowmeter on the front or top panel. The dew point is shown on a touchscreen display. A 4-20mA output is optionally available. Some models have rechargeable battery operation. Please contact Nova for the correct version for your application.



Models:
  • 203K - suitcase cabinet with display on inside; display visible when case is opened
  • 203KX - suitcase cabinet with display on outside; display visible when case is closed
  • 230BT - rectangular bench top style cabinet




 Specifications 

Features

  • Low range of -112° to +68°F (or -80° to +20°C) Dew Point
  • Easy to operate; long-life sensor
  • Fast warm-up and response
  • Use on any non-corrosive industrial gases or air
  • Built-in sample pump, filter, and flow meter
  • Blocking solenoids used to trap low DP gas in cell for quick purge down and to protect cell when not in use, or during initial sample line purge
  • Touch-screen display of gas readings
  • Rechargeable battery operation
  • Suitcase style case with choice of display locations

SPECIFICATIONS

Method of Detection: Aluminum oxide capacitance sensor
Ranges Available: -80° to +20°C or -112° to +68°F
Resolution: 1°
Accuracy and Repeatability:
± 1°C from -60° to +20°C Dew Point; ± 2°C from -80° to -60°C
± 1°F from -76° to +68°F Dew Point; ± 2°F from -112° to -76°F
Drift:
± 0.5°C Dew Point max. for temperatures between -29 to +4°C
± 1°F Dew Point max. for temperatures between -20 to +40°F
Response Time (T-90):
Less than 30 seconds at higher dew points;
Lower dew points may take up to 1 minute (sensor must be in pre-dried state)
Ambient Temperature Range: 0° to 120°F (-18° to 50°C)
Linearity: ± 2°F / C over measurement range
Size and Weight:
K style - approx. 35 x 15 x 17 cm @ 5.5 kg (14" W x 6" H x 6½" D @ 12 lbs)
Power:
115VAC 60 Hz for recharging (230VAC 50 HZ also available)
Output: 4-20mA

(Nova reserves the right to specification changes which may occur with advances in design without prior notice.)