Canada
270 Sherman Avenue N.
Hamilton, ON L8L 6N5
Phone: (905) 545 2003• Fax: (905) 545 4248
United States
1925 Pine Avenue
Niagara Falls, NY 14301
Phone: (716) 285-0418 • Fax: (716) 282-2937
 
Email: sales@nova-gas.com or websales@nova-gas.com
All Categories > Multigas Analyzers > Item # 600 Series  


Item # 600 Series, Portable Multi-Gas Ambient Air Analyzer


For ambient air or confined space monitoring for up to any four choices of O2, CO2, NO, NO2, CO, LEL combustibles, temperature, and % RH. Single gas versions are also available for H2S, SO2, Cl2, HCL & NH3.

 





 Specifications 

Method of Detection

O2, CO, SO2, NO, NO2, H2, CL2, NH3 and H2S by electrochemical sensor, % RH by dielectric ceramic sensor, CO2 by infrared sensor, combustibles by catalytic sensor

Ranges Available

0-100 PPM NO
0-100% RH

0-20 PPM H2S

0-20 PPM NH3

0-20.0 PPM CL2

0-20.0 PPM NO2

0-200 PPM CO

0-200 PPM H2

0-25.0% O2

0-50°C Temp.

0-50.0 PPM SO2

0-5000 PPM CO2

Resolution

0.1% on % gas, 1 PPM on PPM gas

Accuracy & Repeatability

±2% of full scale

Drift

Less than 1% of full scale or 1 PPM per 8 hours, whichever is greater, with constant ambient temperature

Response Time (T-90)

Varies with individual sensor but generally under 2 minutes for full reading

Ambient Temperature Range

4 to 49 ºC
40 to 120
ºF

Linearity

±2% of full scale or 1 PPM, whichever is greater

Size & Weight

15" W x 7" H x 12" D @ 14 lbs. (374 x 175 x 305 mm @ 6.4 kg)

Power

115 VAC, 60 Hz for recharging
220 VAC, 50 Hz

Output Options

0-1 VDC
4-20 mA

Alarms

Audible and visual alarms are optionally available as well as low battery and low flow

Unique Applications

      Warm up time is approximately two minutes however NO and NH3 may take up to 2 hours for full stability. The operating humidity range for this series of analyzers is 0-90% RH non-condensing. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.

Models & Configuration Guidelines

      Standard system comes with 'K' series suitcase-style enclosure, built-in sampling pump, flow meter, pre-filter, liquid block, 12 ft (3.6 m) sample hose, rechargeable battery, and recharger.

Gases Measured

0) NH3 - 0-20 PPM
1) O2 - 0-25.0%

2) CO - 0-200 PPM

3) CO2 - 0-5000 PPM

4) LEL Combustibles (not avail. for 501 Series)

5) H2S - 0-20 PPM

6) SO2 - 0-50 PPM

7) NO - 0-100 PPM

8) NO2 - 0-20 PPM

9) Cl2 - 0-20 PPM

Industry Type

Indoor Air Quality
Personnel Safety

Steel Making



Print Back