All Categories > Multigas Analyzers > Item # 362 Series  


Item # 362 Series, Portable Flue Gas Analyzer for Oxygen, Carbon Monoxide and Oxides of Nitrogen


The Nova 362 Series are portable analyzers that can provide a detailed analysis of flue gas composition. They have been designed to be accurate, reliable, and easy to use and maintain. Model 362 has customer-replaceable sensors, which respond quickly to the O2, CO, and NOx present in the flue gas sample. The sensor life expectancy is between 3-4 years.

In operation, a built in sample pump draws in the flue gas sample through the S.S. probe, 12 ft. sample hose, condensate removal filter, secondary filter, flow meter, and then on to the three sensors. The detected O2, CO, and NO are displayed on separate LCD digital meters, which have a switchable backlight for use in dark areas. The analyzer can be operated for long periods unattended.

The Nova 362 can also optionally indicate net stack temperature for doing fuel efficiency calculations (362T). The temperature sensor is built into the sampling probe. Efficiency charts for each fuel are provided.

A rechargeable 'gel cell' battery provides enough power for about 16 hours of continuous operation and the analyzer can be used while it is being recharged. A red LED indicates when to recharge and a green LED verifies that it is receiving recharging power. The battery recharger is included with the analyzer. The rugged dust-tight and water resistant case is ideal for use in harsh environments.




 Specifications 

Method of Detection

Customer replaceable electrochemical O2, CO and NO sensor

Ranges Available

0-200 to 0-2000 PPM NOx (as NO); 0-1800°F (0-1000°C) net stack temperature (362T)
0-25.0% O2; 0-2000 PPM to 0-4.00% CO (switch selectable)

Resolution

0.1% O2; 1 PPM on CO and NOx

Accuracy & Repeatability

<1% of O2; ± 10 PPM CO; ± 3 PPM NOx

Drift

±2% of full scale per 8 hours of continuous operation

Response Time (T-90)

10-12 seconds for O2; 20-30 seconds for CO and NOx

Ambient Temperature Range

0 to 40 ºC
32 to 105
ºF

Linearity

±1% of full scale of each gas measured

Size & Weight

K Style: 10 1/2" W x 14" L x 6" H @ 13 lbs (27 x 35 x 15 cm @ 6 kg)
WP Style: 18" W x 12" L x 5 1/4" H @ 13 lbs (45 x 30 x 13 cm @ 6 kg)

Power

AC/DC Operation, 115 VAC 60 Hz for recharging (Other voltages available.)

Output Options

4-20 mA or 0-1 VDC

Unique Applications

All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.

Primary Gas to be Measured

Carbon Monoxide (CO)
Oxides of Nitrogen (NOx)

Oxygen (O2)

Industry Type

Flue Gas or Emissions