Method of Detection
|
|
Electrochemical O2 sensor Infrared detector for CO2
|
|
Ranges Available
|
|
0-25% O2 (Other ranges available) 0-3000 PPM; 0-5000 PPM; 0-1%; 0-10%; 0-20%; 0-50%; 0-100% CO2
|
|
Resolution
|
|
0.1% on % versions or 10 PPM on PPM versions
|
|
Accuracy & Repeatability
|
|
Within ±2% full scale
|
|
Drift
|
|
Within 1% full scale per 8 hours of continuous operation
|
|
Response Time (T-90)
|
|
O2 less than 10 seconds; CO2 less than 30-40 seconds to T-90
|
|
Ambient Temperature Range
|
|
4 to 49 ºC 40 to 120 ºF
|
|
Linearity
|
|
±1.0% of full scale
|
|
Size & Weight
|
|
K Style: 14" L x 10 1/2" W x 6" H @ 12 lbs (36 x 27 x 15 cm @ 5.5 kg) WP Style: 10" L x 7 1/2" H x 6 1/2" D @ 8 lbs (25.5 x 19 x 26.6 cm @ 3.6 kg)
|
|
Power
|
|
AC/DC Operation, 115 VAC, 60 Hz for recharging
|
|
Output Options
|
|
0-1 VDC 4-20 mA
|
|
Unique Applications
|
|
All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.
|
|
Primary Gas to be Measured
|
|
Carbon Dioxide (CO2) Oxygen (O2)
|
|
Industry Type
|
|
C/A Room Monitoring
|
|