Canada
270 Sherman Avenue N. Hamilton, ON L8L 6N5
Phone: (905) 545 2003• Fax: (905) 545 4248 |
United States
1925 Pine Avenue Niagara Falls, NY 14301
Phone: (716) 285-0418 • Fax: (716) 282-2937 |
|
Email: sales@nova-gas.com or websales@nova-gas.com
|
|
|
|
|
Item # 5000 Series, Portable Flue Gas Analyzer for Any Combination of O2, CO, CO2, NOx, SO2and Stack Temperature
|
|
|
|
The Nova 5000 Series Portable Flue Gas analyzer has been designed for accuracy, reliability, ease of use and ease of service, providing a detailed analysis of flue gas composition. The 5000 uses customer replaceable sensors which respond quickly to the O2, CO2, SO2, CO, and NOx (as NO) present in the flue gas sample. The O2 sensor life expectancy is typically 3-4 years. The CO, SO2, and NOx sensor life is typically 2-3 years. The infrared CO2 sensor should not need to be replaced under normal conditions of use. In operation, a built in sample pump draws in the flue gas sample through the stainless steel probe, 12ft. sample house, water separating filter, secondary filter, PTFE liquid blocker and flow meter then on to the gas sensors. The detected O2, CO2, SO2, CO, and NOx are displayed on separate large LCD meters with switchable backlights for use in dark areas. Optional with the Nova 5000 Series is a stack temperature reading. A rechargeable 'gel cell' battery provides enough power for about 4-5 hours of continuous operation and the analyzer can be used while it is being recharged. A red LED tells when to recharge and a green LED verifies that it is receiving recharging power. The Model 5000 series case is rugged, dust and water resistant when closed. A built in printer, which also shows date and time, is optionally available. A built in data logger is also available which can be uploaded to a PC.
|
|
|
|
Method of Detection
|
|
Customer replaceable electrochemical O2, CO, SO2, and NO sensors Solid state infrared detector for CO2
|
|
Ranges Available
|
|
0-1800°F (0-1000 °C) net stack temperature 0-20.00% CO2 0-200 or 0-2000 PPM NOx (as NO or NO + NO2) 0-200 or 0-2000 PPM SO2 0-25.0 % O2 0-9999 PPM CO
|
|
Resolution
|
|
±0.1% CO2, 1 PPM, 1°F (1°C)
|
|
Accuracy & Repeatability
|
|
±10 PPM for all gases 0.1%
|
|
Drift
|
|
<2% full scale per 8 hours of continuous operation
|
|
Response Time (T-90)
|
|
< 30 seconds for 90% of O2, CO2, CO and SO2, < 60 seconds for 90% NO
|
|
Ambient Temperature Range
|
|
Operating : 32 to 122°F (0 to 50°C); Sample : Up to 1200°F (635°C) (standard probe)
|
|
Linearity
|
|
±1% of full scale for each gas measured
|
|
Power
|
|
12 V battery operation, 115 VAC, 60 Hz for recharging 220 VAC, 50 Hz
|
|
Output Options
|
|
0-1 V for each gas measured. RS232 or RS485, printer 4-20 mA
|
|
Primary Gas to be Measured
|
|
Carbon Monoxide (CO) Oxygen (O2)
|
|
Industry Type
|
|
Flue Gas or Emissions
|
|
|
|
|
|
|
|
|
|
|
|
|